Test Mode = ACW
Select by
∧
or
∨
Test Mode = DCW
Select by
∧
or
∨
Test Mode = IR
Select by
∧
or
∨
Voltage = XXXXV
Range:500-1000
Test Mode = GND
Select by
∧
or
∨
Current = XX.XXA
Range:3.00-30.00
3. Test Mode Selection
After pressing the "SET" key, the program will enter the test mode selection, and the LCD will
display:
AC withstand voltage test
DC withstand voltage test
or
Insulation resistance test
Ground resistance test
or
or
Please use the "+" or "-" key on the panel to enter the test mode to be set: ACW, DCW, IR or
GND.
4. Output voltage and current setting
After selecting the test mode and pressing the "SET" key, the program will enter the output
voltage setting mode of AC / DC withstand voltage or insulation resistance test or the output
current mode of ground resistance test.
AC withstand voltage test
DC withstand voltage test
Or
Insulation resistance test
Ground resistance test
or
or
Please use the "+" or "-" key on the panel to input the desired output voltage or current.
5. Leakage current or insulation resistance upper limit (HI-Limit) setting
After setting the output voltage and pressing the "SET" key, the program will enter the leakage
current or insulation resistance upper limit setting mode or ground resistance upper limit setting
mode of the AC / DC withstand voltage test. The LCD will display:
AC withstand voltage test
DC withstand voltage test
or
Voltage = XXXXV
Range:0-5000
Voltage = XXXXV
Range:0-6000
HI-Limit = XX.XXmA
Range: 0.01-12.00
HI-Limit = XX.XXmA
Range: 0.01-6.00