SECTION 9 

– Troubleshooting 

9.1  ERROR CODES 

Refer  to  the  table  below  to  diagnose  error  codes.  If  you  require  further  assistance  contact  Jenway 

Help (refer to section 9.2).  

Error Code 

Error Name 

Issue 

Err 1 

Dark cal error 
     
 
 

This  error  occurs  when  the  6305  closes  the  shutter  to  block 
light  entering  the  sample  chamber.  If  the  detector  output  does 
not  fall  to  a  level  normally  associated  with  a  dark  cal  then  this 
error  is  indicated.  The  most  likely  cause  is  that  the  sample 
chamber lid is not closed. 
 

Err 2 

Light cal error 
 
     
 
 
 

This error indicates that there is  insufficient  light  to calibrate to 
100%.  The  most  likely  cause  is  that  the  light  at  the  selected 
wavelength  is  being  absorbed  by  a  sample  in  the  sample 
chamber.  
 
This error can also be caused if the lamp has failed.  
 

Err 3 

Concentration 
calibration error 
     
 

The  calculated  factor  is  outside  the  instrument  range. 
Recalibrate  the  instrument  using  a  standard  of  lower 
concentration.   
 

Err 5 

Lamp Failure 
     
 

The instrument is unable to detect a peak signal for wavelength 
calibration.  Possible  causes  of  this  error  are  that  the  lamp  has 
failed  or  that  a  sample  has  been  left  in  the  sample  chamber 
during start-up. 
 

Err 6 

Power on self-test 
error    
 
 

Detector signal with the lamp off exceeds the preset threshold. 
Possible cause is that the sample chamber lid is not closed. 
 

Err 7 

Power on self-test 
error   
 

The  instrument  is  unable  to  detect  the  monochromator 
reference  position  sensor.  Please  contact  Bibby  Scientific 
service or your local distributor.  

 

9.2  

TECHNICAL SUPPORT 

Jenway  have  a  dedicated  Technical  Support  team  made  up  of  experienced  scientists  who  are  on 

hand to help with any applications advice and questions you may have about our  products  and  how 

to use them. If you require any technical or application assistance please contact the team at: 

E-mail: 

[email protected]

 

Phone: +44 (0)1785 810433 

Содержание 6305

Страница 1: ...Operating Manual 635 015 REV B 06 14 Model 6305 Spectrophotometer...

Страница 2: ...2...

Страница 3: ...val of covers and attempted adjustment or service by unqualified personnel will invalidate any warranty and incur additional charges for repair 7 Reference should always be made to the Health and Safe...

Страница 4: ...n de ces produits chimiques doivent tre respect s soigneusement Lorsqu elles ne sont pas disponibles ces donn es doivent tre demand es dans tous les cas au fournisseur avant de poursuivre 8 Il est imp...

Страница 5: ...eine geeignete Sicherheitsausr stung und pers nliche Schutzausr stung zu verwenden 9 Falls vermutet wird dass der Sicherheitsschutz auf irgendeine Weise beeintr chtigt wurde ist das Spektrophotometer...

Страница 6: ...are o utilizzare lo strumento 1 Gli spettrofotometri della serie 63 sono progettati per essere utilizzati da personale qualificato che conosce i principi e le applicazioni ad essi collegati Per ulteri...

Страница 7: ...tiche di laboratorio quando si manipolano campioni sostanze chimiche reagenti e apparecchiature accessorie per eseguire misurazioni e analisi con questo strumento Attrezzature idonee per la protezione...

Страница 8: ...RICS MEASUREMENTS 18 4 1 TRANSMISSION MEASUREMENT MODE 18 4 2 ABSORBANCE MEASUREMENT MODE 18 SECTION 5 CONCENTRATION MEASUREMENTS 20 5 1 USING A KNOWN STANDARD 20 5 2 USING A KNOWN FACTOR 20 SECTION 6...

Страница 9: ...ontamination Combined with a mechanically rigid structure this model provides a system with fast warm up low drift and high reliability 1 2 INSTRUMENT SPECIFICATION 6305 Wavelength Range 198 to 1000nm...

Страница 10: ...he mains fuse is housed within the POWER IN socket When replacing the fuse the unit must be disconnected from the mains supply The fuse rating is 2A F fast blow type In the event of the fuse failing a...

Страница 11: ...g 2 3 1 Display 1 Primary display area Transmission Absorbance Concentration 2 Primary display adjust annunciator 3 Secondary display area Wavelength Factor 4 Primary display units 5 Secondary display...

Страница 12: ...u option 4 CAL Key Initiates a calibration routine 5 Print Key Provides a printout of the current reading with an incremental sample number When pressed for the first time after a calibration the prin...

Страница 13: ...ess panel and all ventilation slots must NOT be covered or obstructed at any time 2 Power switch On off switch for the unit 3 Power in socket IEC type connection socket for mains cable 4 Output socket...

Страница 14: ...is travelling through The Beer Lambert law states that A is the absorbance is the molar absorption coefficient l mol 1 cm 1 c is the concentration mol l 1 l is the path length cm This law shows that...

Страница 15: ...enon lamp is focused onto the grating with 1200 lines per millimeter which separates the light into discreet wavelengths The diffracted spectrum of light then passes through a further slit and lens ar...

Страница 16: ...to 260nm Below this level quartz cuvettes must be used NOTE the plastic cuvettes supplied with the unit are not suitable for use below a wavelength of 320nm b Plastic disposable cuvettes should only b...

Страница 17: ...In the preparation of samples and standards high grade borosilicate glass and AR grade chemicals and reagents must be used Good quality deionised water or other suitable solvents must be used for diss...

Страница 18: ...screen Subsequent samples can be measured in the same way If the wavelength is adjusted between sample measurements then the instrument must be calibrated again before more samples can be measured 4 2...

Страница 19: ...19 measured...

Страница 20: ...ously Use the arrow keys to change the concentration to the required value for the standard Press the CAL or RETURN key to calibrate the instrument to the standard concentration The instrument will ca...

Страница 21: ...also be sent to the 6305 via the serial interface using Windows HyperTerminal for example ASCII D or d Same as pressing the PRINT key ASCII T CR Outputs transmission and wavelength separated by an AS...

Страница 22: ...readily incorporated into most spreadsheet software packages 6 1 RS232 Output The bi directional RS232 interface is available on the rear panel 25 way D type connector The connections are as follows T...

Страница 23: ...PC GND 7 5 GND Note the interface cable kit 542 009 can be used to implement the above interconnections 6 2 Recorder Output This is available via the 4mm rear panel sockets The level is proportional...

Страница 24: ...or use with computer without a serial port 543 001 40 Column printer supplied with interface cable paper roll and ribbon and universal power supply 060 287 Paper roll for printer quantity 1 060 288 Pr...

Страница 25: ...or commercial use and will give rise to errors if used 8 2 LAMP REPLACEMENT 8 2 1 Xenon Flash Lamp Module Replacement Before replacing the lamp disconnect the unit from the mains supply Access to the...

Страница 26: ...26...

Страница 27: ...e Department for a reference number which you should include with your faulty equipment Please also ensure you include a clear description of the fault and a completed copy of our Decontamination Cert...

Страница 28: ...is outside the instrument range Recalibrate the instrument using a standard of lower concentration Err 5 Lamp Failure The instrument is unable to detect a peak signal for wavelength calibration Possi...

Страница 29: ...29 SECTION 10 Declaration of Conformity...

Страница 30: ...748 Fax 33 1 64 45 13 00 Fax 44 0 1785 810405 e mail info bibby scientific com e mail bsf bibby scientific fr e mail bibby bibby scientificasia com www bibby scientific com www bibby scientific com ww...

Отзывы: