SECTION 4 

– PHOTOMETRICS MEASUREMENTS 

 

The  6305  can  perform  simple  measurements  of  %  transmittance  or  absorbance.  The  sample  is 

measured at one wavelength and at one point in time.  

4.1 

TRANSMISSION MEASUREMENT MODE 

To select % Transmittance use the LEFT or RIGHT arrow keys to move the cursor to the %T indicator 

on the menu bar (refer to figure 2.3.1). The primary display will show the transmittance with %T units. 

To adjust the wavelength use the UP and DOWN arrow keys. 

The  calibration  must  be  performed  at  the  same  wavelength  at  which  the  sample  will  be  measured. 

Insert  a  cuvette  containing  the  blank  (clear)  solution  into  the  sample  chamber  and  close  the 

instrument  lid.  Press  the  CAL  key.  The  instrument  will  perform  a  zero  %  transmission  calibration 

followed by a 100% transmission calibration. An internal shutter is automatically activated to perform 

the  zero  %  setting  and  this  part  of  the  routine  is  therefore  independent  of  the  solution  in  the  light 

path).  

Once the calibration has been performed a sample can be measured. Remove the cuvette containing 

the  blank  solution  and  place  a  cuvette  containing  the  sample  to be  measured  in  the  sample  holder. 

Close the instrument lid and the % transmittance result will be shown on the screen.  

Subsequent  samples  can  be  measured  in  the  same  way.  If  the  wavelength  is  adjusted  between 

sample  measurements  then  the  instrument  must  be  calibrated  again  before  more  samples  can  be 

measured. 

4.2 

ABSORBANCE MEASUREMENT MODE  

To select absorbance use the LEFT or RIGHT arrow keys to move the cursor to the ABS indicator on 

the menu bar (refer to figure 2.3.1). The primary display will show the absorbance with ABS units. To 

adjust the wavelength use the UP and DOWN arrow keys. 

The  calibration  must  be  performed  at  the  same  wavelength  at  which  the  sample  will  be  measured. 

Insert  a  cuvette  containing  the  blank  (clear)  solution  into  the  sample  chamber  and  close  the 

instrument  lid.  Press  the  CAL  key.  The  instrument  will  perform  a  zero  %  transmission  calibration 

followed by  a 0.000 Absorbance calibration. An internal shutter is automatically activated to perform 

the zero % setting and this part of the routine is therefore independent of the solution in the light path.  

Once the calibration has been performed a sample can be measured. Remove the cuvette containing 

the  blank  solution  and  place  a  cuvette  containing  the  sample  to be  measured  in  the  sample  holder. 

Close the instrument lid and the absorbance result will be shown on the screen.  

Subsequent  samples  can  be  measured  in  the  same  way.  If  the  wavelength  is  adjusted  between 

sample  measurements  then  the  instrument  must  be  calibrated  again  before  more  samples  can  be 

Содержание 6305

Страница 1: ...Operating Manual 635 015 REV B 06 14 Model 6305 Spectrophotometer...

Страница 2: ...2...

Страница 3: ...val of covers and attempted adjustment or service by unqualified personnel will invalidate any warranty and incur additional charges for repair 7 Reference should always be made to the Health and Safe...

Страница 4: ...n de ces produits chimiques doivent tre respect s soigneusement Lorsqu elles ne sont pas disponibles ces donn es doivent tre demand es dans tous les cas au fournisseur avant de poursuivre 8 Il est imp...

Страница 5: ...eine geeignete Sicherheitsausr stung und pers nliche Schutzausr stung zu verwenden 9 Falls vermutet wird dass der Sicherheitsschutz auf irgendeine Weise beeintr chtigt wurde ist das Spektrophotometer...

Страница 6: ...are o utilizzare lo strumento 1 Gli spettrofotometri della serie 63 sono progettati per essere utilizzati da personale qualificato che conosce i principi e le applicazioni ad essi collegati Per ulteri...

Страница 7: ...tiche di laboratorio quando si manipolano campioni sostanze chimiche reagenti e apparecchiature accessorie per eseguire misurazioni e analisi con questo strumento Attrezzature idonee per la protezione...

Страница 8: ...RICS MEASUREMENTS 18 4 1 TRANSMISSION MEASUREMENT MODE 18 4 2 ABSORBANCE MEASUREMENT MODE 18 SECTION 5 CONCENTRATION MEASUREMENTS 20 5 1 USING A KNOWN STANDARD 20 5 2 USING A KNOWN FACTOR 20 SECTION 6...

Страница 9: ...ontamination Combined with a mechanically rigid structure this model provides a system with fast warm up low drift and high reliability 1 2 INSTRUMENT SPECIFICATION 6305 Wavelength Range 198 to 1000nm...

Страница 10: ...he mains fuse is housed within the POWER IN socket When replacing the fuse the unit must be disconnected from the mains supply The fuse rating is 2A F fast blow type In the event of the fuse failing a...

Страница 11: ...g 2 3 1 Display 1 Primary display area Transmission Absorbance Concentration 2 Primary display adjust annunciator 3 Secondary display area Wavelength Factor 4 Primary display units 5 Secondary display...

Страница 12: ...u option 4 CAL Key Initiates a calibration routine 5 Print Key Provides a printout of the current reading with an incremental sample number When pressed for the first time after a calibration the prin...

Страница 13: ...ess panel and all ventilation slots must NOT be covered or obstructed at any time 2 Power switch On off switch for the unit 3 Power in socket IEC type connection socket for mains cable 4 Output socket...

Страница 14: ...is travelling through The Beer Lambert law states that A is the absorbance is the molar absorption coefficient l mol 1 cm 1 c is the concentration mol l 1 l is the path length cm This law shows that...

Страница 15: ...enon lamp is focused onto the grating with 1200 lines per millimeter which separates the light into discreet wavelengths The diffracted spectrum of light then passes through a further slit and lens ar...

Страница 16: ...to 260nm Below this level quartz cuvettes must be used NOTE the plastic cuvettes supplied with the unit are not suitable for use below a wavelength of 320nm b Plastic disposable cuvettes should only b...

Страница 17: ...In the preparation of samples and standards high grade borosilicate glass and AR grade chemicals and reagents must be used Good quality deionised water or other suitable solvents must be used for diss...

Страница 18: ...screen Subsequent samples can be measured in the same way If the wavelength is adjusted between sample measurements then the instrument must be calibrated again before more samples can be measured 4 2...

Страница 19: ...19 measured...

Страница 20: ...ously Use the arrow keys to change the concentration to the required value for the standard Press the CAL or RETURN key to calibrate the instrument to the standard concentration The instrument will ca...

Страница 21: ...also be sent to the 6305 via the serial interface using Windows HyperTerminal for example ASCII D or d Same as pressing the PRINT key ASCII T CR Outputs transmission and wavelength separated by an AS...

Страница 22: ...readily incorporated into most spreadsheet software packages 6 1 RS232 Output The bi directional RS232 interface is available on the rear panel 25 way D type connector The connections are as follows T...

Страница 23: ...PC GND 7 5 GND Note the interface cable kit 542 009 can be used to implement the above interconnections 6 2 Recorder Output This is available via the 4mm rear panel sockets The level is proportional...

Страница 24: ...or use with computer without a serial port 543 001 40 Column printer supplied with interface cable paper roll and ribbon and universal power supply 060 287 Paper roll for printer quantity 1 060 288 Pr...

Страница 25: ...or commercial use and will give rise to errors if used 8 2 LAMP REPLACEMENT 8 2 1 Xenon Flash Lamp Module Replacement Before replacing the lamp disconnect the unit from the mains supply Access to the...

Страница 26: ...26...

Страница 27: ...e Department for a reference number which you should include with your faulty equipment Please also ensure you include a clear description of the fault and a completed copy of our Decontamination Cert...

Страница 28: ...is outside the instrument range Recalibrate the instrument using a standard of lower concentration Err 5 Lamp Failure The instrument is unable to detect a peak signal for wavelength calibration Possi...

Страница 29: ...29 SECTION 10 Declaration of Conformity...

Страница 30: ...748 Fax 33 1 64 45 13 00 Fax 44 0 1785 810405 e mail info bibby scientific com e mail bsf bibby scientific fr e mail bibby bibby scientificasia com www bibby scientific com www bibby scientific com ww...

Отзывы: