31
V-530 measurement parameters
V-550/560/570
measurement
parameters
Photometric mode Sample
Photometric mode
Sample
Response
Slow
Response
Slow
Wavelength
250 nm
Band Width
2.0 nm
Start Time
0 sec
Wavelength
250 nm
End Time
60 sec
Start Time
0 sec
Data Pitch
5.00 sec
End Time
60 sec
Data Pitch
5.00 sec
PMT Voltage
300 V
Note: PMT voltage
varies
with the instrument. Set accordingly.
CAUTION:
Do not loosen screws other than the adjustment screw for the stopper.
(3)
Turn the D
2
lamp stopper adjustment screw (Fig 4.9 or Fig, 4.10) and search for a
position where the data value (value displayed on the CRT) is maximum.
(4)
Replace the inner lid and outer lid of the light source unit.
4.5.2.3
Adjustment of iRM type light source mirror
When the light source is replaced, if the beam from the lamp is not correctly incident
upon the entrance slit, troubles will arise, such as lack of luminous energy, bend of the
baseline, and deviation of the measurement value. Whenever the lamp is replaced, be
sure to adjust the light source mirror.
WARNINGS
:
(1) The lamp is hot when lit. Do not touch with bare hands.
(2) Wear glasses (ordinary glasses or sunglasses) during adjustment to
protect your eyes from ultraviolet rays.
(3) Do not touch the wiring either with your hand or with a screwdriver.
Contact with the wiring may cause electric shock.
CAUTION:
Loosen only the screws required for adjustment.
<<
Adjustment of light source mirror for halogen lamp
>>
(1)
Turn on the instrument.
(2)
Select [Envir.] from the main menu to display the [Environment] menu.
(3)
Select [Single Beam] from the [Environment] menu to display the single beam
measurement screen. (Refer to Section 9-5 of the Operations manual.)
(4)
Display the parameter screen and set parameters as shown below.