22
Figure 4.4 D
2
lamp spectra with an emission line at 656.1 nm
(5)
The wavelength repeatability is normal if it is within the following criterion.
Criterion: Within
±
0.1 nm of the average value of the peaks of the spectrum with
an emission line at 656.1 nm
(6)
Display the [Environment] menu again and return to the original settings (i.e. check
the [Deuterium] and [Halogen] boxes).
4.3.3 Wavelength Accuracy
To check wavelength accuracy, measure the spectra of the D
2
lamp with an emission
line at 656.1 nm and 486.0 nm.
(1)
Light only the Deuterium lamp (see item (1) in Section 4.3.2).
(2)
Start [Spectrum Measurement] from the [Spectra Manager] window and set the
following parameters.
<<
Parameters for spectrum with an emission line at 656.1 nm
>>
V-530 measurement parameters
V-550/560/570
measurement
parameters
Photometric Mode Sample
Photometric Mode Sample
Response
Fast
Response
Fast
Scanning Speed
40 nm/min
Band Width
0.5 nm
Start
658 nm
Scanning Speed
10 nm/min
End
654 nm
Start
658 nm
Data Pitch
0.1 nm
End
654 nm
Display
0~100
Data Pitch
0.025 nm
No. of Cycles
1
Display
0~100
PMT Voltage
300 V
No. of Cycles
1
Note: PMT voltage
varies
with the instrument. Set properly.
<<
Parameters for spectrum with an emission line 486.0 nm
>>
Start
488 nm
End
484 nm
All other parameter settings are the same parameters as shown above.