Wavelength scanning
10, 20, 40, 100, 200, 400, 1000, 2000, 4000
nm/min
Wavelength moving speed
8000 nm/min
Data pitch
0.025, 0.05, 0.1, 0.2, 0.5, 1, 2, 5,10 nm/data
(spectrum measurement)
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20 sec/data
(time course measurement)
Baseline stability
±
0.0004 Abs/hour
(value obtained more than one hour after
turning ON the power when temperature
variation is within 5
°
C, wavelength: 250 nm,
response: Slow, and spectral bandwidth: 2nm.)
Baseline flatness
±
0.001 Abs
(value obtained after baseline correction when
temperature
variation
is
within
5
°
C,
wavelength: 200 to 850 nm, response: Medium,
spectral bandwidth: 2 nm, and wavelength
scanning: 100 nm/min)
Detector
Photomultiplier tube
Power requirements
100, 115, 200, 220, 230, 240 V
±
10 %, 130 W
Dimensions and weight
460(W)
×
595(D)
×
260(H)mm (excluding
protrusions)
30.5 kg
Note: Photometric repeatability, photometric accuracy and baseline flatness are the
values obtained more than one hour after the “Power” switch is turned ON.
1.3
V-560 UV/VIS Spectrophotometer Specifications
Optical system
Double monochromator
UV/VIS region: 600 lines/mm plane grating
Czerny-Turner mount
Double beam type
Resolution
0.1 nm
Light source
Deuterium lamp: 190 to 350 nm
Halogen lamp: 330 to 900 nm
Light
source
changeover
wavelength
Any wavelength between 330 and 350 nm can
be selected.
Wavelength range
190 to 900 nm
Wavelength repeatability
±
0.1 nm (at a spectral bandwidth of 0.5 nm)
Wavelength accuracy
±
0.3 nm (at a spectral bandwidth of 0.5 nm)
Spectral bandwidth
0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray-light mode)
Wavelength display
In increments of 0.1 nm
Photometric mode
Abs, %T. %R, Sample, Reference
Photometric range
-2 to 5 Abs
Photometric display
0.0 to 200.0 %T
-2 to 5 Abs