20
Note: This data is used for correcting the baseline spectrum (0 Abs line) to be
measured in (4) below.
(4)
Measure baseline spectrum (0 Abs line) as follows. Click the <Start> button to
measure the baseline spectrum (0 Abs line) using the above parameters with the
sample chamber empty. A measurement example is shown in Fig 4.1.
(5)
The baseline flatness is normal if the measurement result is within the following
criterion.
<V-530 criterion>
Within
±
0.001 Abs (1000 ~ 200 nm)
<V-550/560/570
criterion>
Within
±
0.001 Abs (850 ~ 200 nm)
Figure 4.1 Baseline flatness (V-550/560/570)
4.3.2 Wavelength Repeatability
To check wavelength repeatability, measure the spectrum of the Deuterium (D
2
) lamp
with an emission line at 656.1 nm three times repeatedly.
(1)
Light the D
2
lamp as follows.
a) Start [Environment] from the [Spectra Manager] window. The [Environment]
dialog box appears.
Figure 4.2 [Environment] dialog box
b) Select [Hardware Setting] and then click the <Execute...> button. The following
dialog box appears.