(10) Total Flow Value Add adjust
- In FACTORY MODE menu, enter number
"18"
on PC keyboard and press Enter.
- Total value can be inputted. Input below -30000 ~ +30000 at once.
- Total value can be inputted below -30000 ~ +30000, and if it is above this range, enter
continuously.
Ex : If the input value is +80000, enter it 3 times by +30000.
(11) Actual Zero Value Add adjust
- In FACTORY MODE menu, enter number
"19"
on PC keyboard and press Enter.
- Actual Zero Value can be inputted. Input below -30000 ~ +30000 at once.
Total value can be inputted below -30000 ~ +30000, and if it is above this range, enter
continuously.
Ex : If the input value is +80000, enter it 3 times by +30000.