Xonic 5L
19/23
(3) SoundSpeed Offset correction
- In FACTORY MODE menu, enter number
"11"
on PC keyboard and press Enter.
- Check ultrasonic sound speed table according to water temperature and enter the value in
numbers.
- Ex : If water temperature is 25
℃
and sound speed is 1496.7m/s, enter sound speed as a
number 1496.7m/s * 10 =
”1
4967
”
- Check temperature in Display Information
(4) Forward KC Factor correction
- In FACTORY MODE menu, enter number
"12"
on PC keyboard and press Enter.
- Enter error value of measured Q3 flow.
- Ex : If error of measured Q3 flow is -5.5%, enter it as a number
-5.5% * 10 =
”5
5
”