Xonic 5L
21/23
(7) Reverse DeltaT Offset correction
- In FACTORY MODE menu, enter number
"15"
on PC keyboard and press Enter.
- Enter error value of measured Q2 flow.
- When 80mm measurement error is ±0.1 : Enter value ±12ps / When 80mm measurement
error is ±2.5 : Enter value ±300ps
- When 100mm measurement error is ±0.1 : Enter value ±15ps / When 100mm measurement
error is ±2.5 : Enter value ±375ps
- When 150mm measurement error is ±0.1 : Enter value ±18ps / When 150mm measurement
error is ±2.5 : Enter value ±450ps
- When 200mm measurement error is ±0.1 : Enter value ±22ps / When 200mm measurement
error is ±2.5 : Enter value ±550ps