background image

JADEVER SCALE 

SNUG III                                                                                                                               

10

 

 

 

 

. Weight calibration 

 

5-1 List of calibration weights

 (Please get the standard weighs ready before calibration 

pursuant to the stipulated weight) 

 

MODEL 

SNUG III-150  SNUG III-300

SNUG 

III-600 

SNUG 

III-1500 

SNUG 

III-3000 

on1 

50g 100g 

200g 

500g 1000g 

on2 

100g 200g 400g 

1000g 2000g 

on3 

150g 300g 600g 

1500g 3000g 

 

 

5-2 Single-point calibration of weight 

 

 

Specification 

Display on screen 

Procedure 

Under Stop mode, first hold down button

, then 

press down button 

 once to energize the 

scale, and then release button

C o u n t

 

Procedure 

Press down button 

  to enter calibration mode. 

C A L

 

Procedure 

First make sure that nothing is on the scale pan, and 

then press down button 

 to enter zero 

calibration. 

Z ero

 

Procedure 

Perform calibration after on1 is displayed on screen. 

O n 1

 

Procedure 

Press down button 

 to select weights; on1 

indicates 1/3 of full capacity, on2 indicates 2/3 of that, 
and on3 indicates full capacity. Place the selected 
weight on the pan promptly after completing 
selection. 

O n 1

 

O n 2

 

O n 3

 

Содержание SNUGIII Series

Страница 1: ...2 Specificatin of counting function 3 3 Specification of percentage function 3 4 Setting of time and date 4 4 5 5 Setting of various parameters 4 1 Entry and parameter adjustment mode 4 2 Specificati...

Страница 2: ...ects enter the scale Be sure to select the suitable transformer whose output is DC 9V 400ma please use a separate power socket and the transformer delivered by the scale manufacturer in order to avoid...

Страница 3: ...sted of 5 components and they must be assembled together before being used 2 First assemble four sides of windshield A B by inserting A into B from top to bottom or from bottom to up 3 Place the assem...

Страница 4: ...on specification of buttons and display window Button for Start Stop Button for zeroing It returns the weight to zero may return the value displayed on screen to zero position zeroing range is 2 of th...

Страница 5: ...Press down button to shift the weighing units Caution the symbol of weighing unit will flash when the weight becomes unstable during the weighing 3 2 Specification of counting function 1 Under the wei...

Страница 6: ...ntity place the article whose quantity is the same with the selected one on the scale pan and press down button to begin sampling when the sound bi is heard and SE is disappeared the sampling process...

Страница 7: ...then press down button to select Internal option Procedure 3 After completing the modification of internal options press down button to exit in case the next option should be set press down button to...

Страница 8: ...ed may be cancelled temporarily Press down button to select ON OFF and complete the setting Ozt TROY OUNCE ON in use or OFF not in use Dwt PENNY WEIGHT ON in use or OFF not in use MM JPN ON in use or...

Страница 9: ...ime AU 60 Auto Stop 60 min later 2400 4800 Setting of RS 232 transmission function Note Selection of transmission speed 9600 d 0 No limitation of zero display d 1 In case the indication is below zero...

Страница 10: ...ff Auto backlight off No backlight Backlight mode of display window Note there are three backlight modes that may increase effectively the convenience of application on Backlight keeps ON Pr co Signal...

Страница 11: ...rocedure 1 Under Stop mode first hold down button then press down button once to energize the scale and then release button Count Procedure 2 Press down button to enter calibration mode CAL Procedure...

Страница 12: ...cale and then release all pressed buttons L CAL Procedure 2 After clearing away the scale pan press down button to enter zero calibration On0 Procedure 3 After the sound bi is heard place the weight e...

Страница 13: ...right of SNUG III pin 2 is for signal output and pin 5 is for grounding and other pins are in reserve BAUD RATE 2400 OR 4800 OR 9600 bps DATA BIT 8 PARITY BIT N NONE STOP BIT 1 CODE ASCII BIT FORMAT L...

Страница 14: ...ial transmission speed should be set correspondingly USB peripheral devices should be used as follows 1 Connect this electronic scale to a computer with USB line and then start the electronic scale in...

Страница 15: ...t Err5 The weight of an article exceeds the maximal capacity 9e Remove the over heavy article from the pan promptly and avoid the similar error from then to avoid the possible damage Err6 The weight f...

Страница 16: ...we beg your pardon for this 1 You fail to show the warranty of products or the writing of the warranty is too illegible to be discerned 2 The damage of scale is due to the abnormal or incorrect appli...

Отзывы: