JIF-2002 WEIGHING INDICATOR&CONTROLLER
7
2-3 Connecting the Load Cell
※
Do not plug in your power cable until you have completely connected the
load cell.
Screw
Signal
1
Positive Excitation Voltage, (EXC+)
2
Positive Sense Voltage, (SEN+)
3
Negative Sense Voltage, (SEN-)
4
Negative Excitation Voltage, (EXC-)
5
Positive Signal Voltage, (SIG+)
6
Negative Signal Voltage, (SIG-)
7 Shield,
(SHD)
L
To connect your load cell to the weighing Indicator use a six-wire cable
with shield-connect the wires as indicated above. If the JIF-2002 is located
near the Load Cells (Within five meters or a few yards) you may use a
4-wire cable with shield, but first connect screws 1&2 and 3&4 with
independent jumper leads.
L
The analogue output from the Load Cell and input/output signals are
sensitive to electrical noise. Do not bind these cables together as it could
result in cross-talk interface. Please also keep them away from AC power
cables.
※
It could be dangerous by using improper battery or wrong connection of
battery.