JIF-2002 WEIGHING INDICATOR&CONTROLLER
31
2. Select Digital Calibration
An easy way to make calibration by inputting Load Cell’s Full Scale Output
voltage
STEP 1:
Turn the Power Switch OFF on the rear panel. Slide the
SET
switch to the set side
.
STEP 2:
Turn the power switch ON. The screen will show blinking
SELECT.
STEP 3:
Please press
GROSS/NET
key and a blinking
CAL
will show
on the screen. Then press the
↵
key. The screen will show
F-CAL
. Please use the
or
key to choose digital Calibration
(
d-CAL
).
STEP 4:
The screen will show
F-CAL
. Please use the
or
key to
choose digital Calibration (
d-CAL
). Please press the
↵
key.
﹝
1
﹞
Setting Minimum Division
The display of
di 1
shows the smallest division. Use
the
or
key to move through the available divisions.
﹝
1
、
2
、
5
、
10
、
20
、
50
﹞
. Press the
↵
key to set the minimum division.
﹝
2
﹞
Setting Decimal
(
F000 will change---see 4-2
)
The screen will show dp
→
d000.000. A blinking decimal will
show on the screen. Use the
or
key to move through the
available decimal
Point
position. Press the
↵
key to set
the decimal position.
﹝
3
﹞
Setting Maximum Capacity
When setting maximum capacity, the screen will show
CAP
→
C000000
.Use the
or
key to move through digits. Use
Содержание JIF-2002
Страница 1: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 1 CHAPTER 1 INTRODUCTION ...
Страница 5: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 5 CHAPTER 2 INSTALLATION ...
Страница 9: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 9 CHAPTER 3 SPECIFICATIONS ...
Страница 17: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 17 CHAPTER 4 SYSTEMFUNCTIONS ...
Страница 41: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 41 CHAPTER 5 SET POINTS ...
Страница 54: ...JIF 2002 WEIGHING INDICATOR CONTROLLER 54 CHAPTER 6 OPTIONS ...