JIF-2002 WEIGHING INDICATOR&CONTROLLER
37
STEP 3:
Please press
ESC
key and a blinking
INIT
will show on the
screen. Please enter
↵
key.
STEP 4:
Use the or
key to choose
NO
or
YES
. If
NO
is entered, the
screen will show
END
. If
YES
is entered, the screen will
show ……
→
END
.
STEP 5 :
Slide the
SET
switch to the original side for finishing the step.
◎
Calibration Errors
C.Err 1
:
The resolution exceeds 1
:
16,000.
Change the minimum division and maximum capacity within
1
∕
16,000.Resolution ratio= Minimum division
∕
maximum capacity
C.Err 2
:
The load cell output is too large at ZERO calibration.
Add an additional resistor (50k
Ω∼
500K
Ω
) between EXC+
and SIG
—
.
※
Refer to the Right Figure
C.Err 3
:
The load cell output is too small at ZERO calibration.
Add an additional resistor (50k
Ω∼
500K
Ω
) between
EXC+ and SIG+.
※
Refer to the Right Figure