44
- Check that the water temperature is not too high.
- The “taratura sensore H
2
O” (H
2
O sensor setting) value may be too
low. Contact an authorized Jacuzzi® service centre.
X
The display indicates problems with the measurement
unit
- Check that the measurement unit is clean (refer to the section,
“Cleaning the measurement unit”).
- The spotlight (P) and/or (F) photosensor may be damaged. Con-
tact an authorized Jacuzzi® service centre.
F
P
X
The measurement data archive is empty.
- Check that the “step” parameter is not equal to 0 (refer also to sec-
tion, “Displaying and back-up of the measurement values”).
Structure of the menus
and the relative factory settings
1 - SETUP SYSTEM
DATE/HOUR SYSTEM
DAY/MONTH/ YEAR/ HOUR/ MINUTE
LANGUAGE
ENGLISH
PASSWORD
00000
DISPLAY
CONTRAST
88
BACK
LIGHTEN.
NO
SETUP RELAY
SETUP RELAY ALARM
ALARM
TIME
CL
0000
ALARM
TIME
PH
0000
DISAB.SET
POINT
SI
2 - SETUP MEASURES
SELECT VALUE
SELECT CL2
SETUP CL2
RELAY OUTPUTS
SET POINT 1
ON
0.80 ppm
OFF
1.10 ppm
ON TIME
+000
OFF TIME
+000
MODE
RELAY
MODE RELAY
pwm
PWM PERIOD
30 sec
SET
POINT
2
ON
0.80 ppm
OFF
1.10 ppm
ON TIME
+000
OFF TIME
+000
MODE
RELAY
MODE RELAY
pwm
PWM PERIOD
30 sec
CURRENT OUTPUT
OUTPUT
RANGE
4-20 mA
LOW LIMIT
0.00 ppm
HIGH LIMIT
5.00 ppm
CALIBRATION
CLHLORINE
AUTOMATIC
DEFAULT RESET
ANALYSIS LAG
LAG
10 min
SERVICE MENU
CHLORINE GAIN
4.00 ppm
CELL ALARM
on
WATER ALARM
on
SETUP ANALYSIS
TIME METHOD
WASHING TIME
35 sec
ZERO
TIME
10 sec
STABIL.MEASURE
20 sec
AGITATION
TIME
2 sec
REAGENT TIME
3 sec
REAGENT SETUP
REAG. ALARM
off
SETUP RELAY aux1
3 min
OTHER SETTINGS
DEFAULT VALUES
MEASURE
COUNTER
RESET COUNTER
MEASURE OUT SET
SI
MAX.ZERO
MEAS.
10
SELECT
TEMP.
SETUP
TEMP.
CURRENT OUTPUT
OUTPUT RANGE
4-20 mA
LOW LIMIT
0.00 ppm
HIGH LIMIT
5.00 ppm
Содержание booster KIT spa-pack
Страница 2: ......
Страница 11: ...11 spa pack booster spa pack spa spa Jacuzzi Jacuzzi JacuzziEuropeS p A Jacuzzi...
Страница 21: ...21 a1 a a1 a 1 2 b1 b b1 b b b1 a a1 1 2 tubo flessibile hose tuyau flexible Schlauch tubo flexible 3 4...
Страница 73: ...73 T UNI 10637 T 7 pH spa pH pH pH 7 0 pH 4 0 spa pack Q pH pH Jacuzzi Q Q 10 Q 10...
Страница 74: ...74 Q spa pack Q spa pack Jacuzzi spa pack X X spa pack spa pack spa pack X H2 O Jacuzzi Jacuzzi X...
Страница 78: ......
Страница 79: ......