38
Start calibration operation.
Push the Enter key after setting the number of
strokes. The pump starts the countdown.
The pump starts to run for the set number of
strokes as it comes to zero.
Measure liquid volume in the calibration cylinder again.
Enter how much liquid has reduced.
Use the UP and DOWN keys to set the volume reduction.
Push the Enter key once. The screen shows flow volume
per shot.
The screen shows "ERROR! Volume Out Of Range!!" if
the reduction is too little or too large. Enter correct liquid
volume or recalibrate it.
NOTE
Calibration with high viscosity liquid at a full suction
speed will often give rise to this error. Reduce a suction
speed along with viscosity.
Push the ESC key to return to the wait mode.
Pump
stops
5
6
Pump
starts
7
Perform a calibration