73
8. What If ...
1912
yes
Data backup – recover
The new software includes program groups which were not yet available at
the time of the data backup. Please check the individual program groups.
The individual program groups are no longer valid.
1920
yes
Data backup – recover
The data backup could not be performed. Please install a newer software
version.
2754
yes
Error self-test
An error occurred while checking the infrared camera
2755
yes
Error self-test
An error occurred while checking the infrared camera
2756
yes
Error self-test
An error occurred while checking the infrared camera
2757
yes
Error self-test
An error occurred while checking the infrared camera
2758
yes
Error self-test
An error occurred while checking the infrared camera
2759
yes
Error self-test
An error occurred while checking the infrared camera
2760
**
yes
Error IRT
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. Check if the protective foil has been removed from the sensor or if
the sensor has been contaminated. Check if a correct firing tray is used.
The height of the firing tray must correspond with the height of the firing
trays recommended for the respective materials by Ivoclar Vivadent AG.
2761
**
yes
Error IRT
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. The object was not sufficiently dried in the specified time. Please
repeat the procedure after the object has cooled down.
2762
**
yes
Error IRT
An error occurred while trying to identify the object. Check if interfering
heat sources are in the visual range of the infrared camera. Sources of
interference are for example: other furnaces, air conditioning systems,
heaters, light bulbs, Bunsen burners, etc. Do not move the objects after the
start of a program.
* Furnace head opens if this error occurs!
** A program in progress is aborted
*** Error cannot be acknowledged; programs cannot be started.
8.2 Additional error messages
Please contact the Ivoclar Vivadent After Sales Service, if one of the following error numbers is being displayed:
25, 29, 43, 44, 45, 46, 47, 48, 56
103, 107, 108, 109, 143, 144, 145, 146, 147
148, 700, 701, 703, 704, 707, 1010, 1011, 1012
1013, 1014, 1015, 1016, 1017, 1018, 1019, 1024, 1025
1026, 1028, 1143, 1144, 1145, 1146, 1147, 1148, 1207
1300, 1301, 1303, 1304, 1305, 1401, 1402, 1500, 1750
1751, 1752, 1753, 2750, 2751, 2752, 2753, 2770