65

Step 3: 
Load the furnace 

Open the furnace head with the OPEN FURNACE HEAD key and 
place the object on the firing tray in the furnace.

   With the Infrared Technology active, position the objects 

in such a way that the infrared camera has a wide view of 
the objects. Use the markings on the frame plate (see 
 Figure) for orientation purposes.

Step 4: 
Start the program

Press the START key to start the program. The LED lights up in 
green. You can observe the course of the program on the firing 
curve display.

The IRT key (see Chapter 5.1.2) is used to switch to the view of the infrared camera during the closing and predrying 
process. The corresponding key is pressed again to return to the firing curve display.

Once the object has reached the ideal removal temperature, the 
display shows a corresponding notification and the cooling time 
elapsed.  

As soon as the OSD lights up in green, the furnace is ready for the 
next program start.

The furnace head opens automatically at the end of the program. If 
IRT was activated, the IRT2 function is started during cooling. The 
function measures the temperature of the fired objects and shows it 
on the right next to the IRT image.

6. Practical Use

Содержание 747826BU

Страница 1: ...Programat P510 G2 Operating Instructions Valid from Software V7 0...

Страница 2: ...2 Declaration of conformity...

Страница 3: ...n to the operation 23 5 2 Firing programs and programming options 26 5 3 Managing programs 38 5 4 Advanced functions of the furnace Software update 42 5 5 Multimedia functions 56 5 6 Digital Shade Ass...

Страница 4: ...Status Display OSD 13 Operating unit fixture 14 USB connections 15 Card reader 2 Cooling tray 4 Audio ports in out 3 Housing base 16 Cover for head opening mechanism 17 Air vents base 18 USB connectio...

Страница 5: ...g muffle 26 Furnace head sealing ring 28 Sealing surface 27 Infrared camera 38 Screw for connection cover 39 Air vents rear panel 40 Heating element fuse 41 Vacuum hose connection 42 Vacuum hose 29 On...

Страница 6: ...uple plug socket 53 Furnace head vacuum connection 43 Furnace head mounting 44 Furnace head release 49 Heater plug 50 Heater plug socket List of parts 60 LAN download cable 61 Programat Firing Tray Ki...

Страница 7: ...uctions The signs and symbols in these Operating Instructions facilitate the finding of important points and have the following meanings Symbol Note Risks and dangers Important information Contraindic...

Страница 8: ...voltage range shown on the images e g rating plate may differ depending on the voltage version of your furnace 1 5 Notes on the images in the Operating Instructions All images and illustrations in th...

Страница 9: ...ions to assure proper use of the furnace The instructions regulations and notes in these Operating Instructions must be observed The instructions regulations and notes in the material s Instructions f...

Страница 10: ...damaged There is a risk of electric shock upon contact with the heating wire Avoid damage of the insulation by contact with the investment tongs or firing tongs Contraindication Firing trays must not...

Страница 11: ...ntains ceramic fibres and may release fibre dust Do not use compressed air or blow on the furnace thus distributing the dust in the environment and observe the additional notes on page 13 Risk of crus...

Страница 12: ...soft cloth Do not use any solvents Disconnect power before cleaning and allow the furnace to cool down The furnace must be cool before it is packed for transportation purposes Use original packaging f...

Страница 13: ...t Fibre dust has proved to be carcinogenic in animal experiments The heating muffle may only be dismounted by a qualified After Sales Service Centre Information regarding the Safety Data Sheet is also...

Страница 14: ...P510 G2 consists of the following components Furnace base with electronic controls Furnace head with firing chamber Cooling tray Firing table Power cord and hose for vacuum pump 3 2 Hazardous areas a...

Страница 15: ...sing contact your local Ivoclar Vivadent Service Centre We recommend keeping the original packaging for future service and transportation purposes 4 2 Selecting the location Place the furnace on a fla...

Страница 16: ...he two fastening screws 1 of the cooling tray and the transport protection of the infrared camera Secure the cooling tray 2 with the two screws 1 Step 2 Mounting the firing plate Place the firing plat...

Страница 17: ...cables of the furnace head with the furnace base Proceed as follows Connect the vacuum hose 53 Insert the thermocouple plug 51 Make sure that the polarity of the plug is correct Insert the heater plu...

Страница 18: ...mum length of the vacuum hose is 1 6 m Step 6 Mounting the connection cover Mount the connection cover 37 and secure it with the fastening screw 38 Make sure that the lateral catch springs snap into p...

Страница 19: ...uum hose 53 7 Press the leaf spring 44 with a finger lift off the furnace head at the same time and remove it Make sure the furnace head has completely cooled down before it is removed fire hazard 4 5...

Страница 20: ...e Next button is used to reach the next entry screen Step 1 Select the language The first setting is the language selections The touch buttons display keys can be operated by tapping the display Selec...

Страница 21: ...tion about the privacy policy can be found at http www ivoclarvivadent com digitalservices The Next button is used to reach the next entry screen Step 5 Select the date format Confirm the entry with t...

Страница 22: ...power supply voltage is in the acceptable range The power supply is outside the acceptable range Software version The currently installed software version is displayed 4 Installation and Initial Start...

Страница 23: ...ction Home Switch to home screen main menu Program key Press once Display of the currently selected program Press twice Switch to the program selection via number entry IRT key With this key the infra...

Страница 24: ...ot in the acceptable range Cancel entry The entry can be cancelled with this button changed values are not saved Program parameter buttons Pressing these buttons allows you to change program parameter...

Страница 25: ...completed To inform the user that the automatic self test has been successfully completed a pre set not modifiable melody is played Furnace head open and temperature below 560 C To inform the user tha...

Страница 26: ...or Ivoclar Vivadent materials b 500 free individually adjustable programs b 500 free individually adjustable programs on a USB stick The program types are subdivided into program groups Each group con...

Страница 27: ...s in free programs are NOT reset to the factory settings or changed 5 2 2 Program selection The program selection requires only few steps 1 Select program type 3 Select the program 4 Start program or...

Страница 28: ...nged The following information is displayed 1 Information bar Program designation Current furnace temperature 2 Firing curve Closing time holding time Temperature increase rate holding temperature lon...

Страница 29: ...he Options button 2 Press the Deactivate write protection button 3 Press the Close button to leave the Options menu 5 Operation and Configuration Changing program options Pressing the Options button o...

Страница 30: ...ture button 3 Enter the desired predrying temperature and confirm with the green button 4 The predrying temperature was successfully changed Press the Close button to leave the Options menu The progra...

Страница 31: ...of the parameters The furnace is equipped with an automatic plausibility check function The parameters are checked upon each program start In case of contradictory parameter combinations the program s...

Страница 32: ...ong term cooling Value range C OFF or 1 1200 C Value range F 0 or 34 2192 F Long term cooling If long term cooling is activated the furnace cools to the set temperature L at the end of the holding tim...

Страница 33: ...d to a stand by temperature of 403 C in the factory The temperature can be individually set for each program Value range C 100 700 C Value range F 212 1292 F Two stage programs If this function is act...

Страница 34: ...ter the pre vacuum time is over The value V1 is ignored upon the start of a program with individually activated pre vacuum The vacuum remains switched on until V2 is reached V2 must be higher than the...

Страница 35: ...operating mode is the preferred and recommended operating mode for standard veneer and all ceramic firing processes If the IRT operating mode is selected the infrared camera controls the predrying an...

Страница 36: ...the form of a firing curve The estimated remaining time is also displayed in 10 second intervals If no vacuum is activated the vacuum indicator and all the corresponding parameters are blank The proc...

Страница 37: ...f the Display button is pressed while the firing curve is displayed the remaining time will be shown The remaining time indicator informs the user about the time that is left until the process is comp...

Страница 38: ...paused the display shows a flashing Pause and the indicator changes back to the program screen If a program is stopped prematurely Vacuum release is shown in the display during the flooding of the fi...

Страница 39: ...ocess 5 3 1 Copy programs With the copy wizard individual programs program groups and program sectors can be copied The source and target of the copy process can be selected in the copy assistant Prog...

Страница 40: ...he product brand for the current group can be selected 1 Press the corresponding button in the Admin menu 2 Select the desired product brand Confirm the entry with the green button 5 3 2 Reset program...

Страница 41: ...can be renamed only possible for individual programs 1 Press the corresponding button in the Admin menu and select whether the program name or group name should be changed 2 Enter the desired program...

Страница 42: ...lay 4 Select the desired display brightness in percent and confirm the entry with the green button or cancel the entry with the red button The setting was changed To return to the home screen press ei...

Страница 43: ...time Setting options hh mm ss Date Set the current date Setting options according to the set date format Date format Set the date format Setting options dd mm yyyy mm dd yyyy Time zone Set the respec...

Страница 44: ...tion interval Set the notification as to when the next calibration should be conducted Setting options 1 3 6 12 months Resetting heating muffle firing hours to zero If this function is executed the he...

Страница 45: ...tion of the furnace Operating hours Number of operating hours 5 Operation and Configuration Testing push notifications Here push notifications for communicating with the Programat app can be tested Th...

Страница 46: ...utorials Display of the QR code to access the video tutorials 5 4 3 Temperature calibration The sheathed thermocouple and heating muffle of the furnace may be subject to changes which affect the furna...

Страница 47: ...ation failed The calibration value is the difference between the measured current temperature and the desired set temperature At the end of the program open the furnace head and carefully remove the A...

Страница 48: ...nction individual programs and settings can be backed up on a USB flash drive We recommend using this feature e g before a software update or before sending in the furnace for maintenance purposes Mor...

Страница 49: ...modified Ivoclar Vivadent programs may be overwritten during a software update Individual programs are not affected and will not be overwritten 5 4 5 Software Update Software updates can be easily in...

Страница 50: ...bject in the chamber e g firing tray may influence the test result Conduct the heater test immediately after switching on the furnace and before any actual firing procedures are conducted If the furna...

Страница 51: ...agnostic button in the Diagnosis menu 2 Generate a diagnostic file Connect a USB stick with the furnace Press the Execute button 3 Once the diagnostic file has been generated one of the following mess...

Страница 52: ...of water in the insulation of the firing chamber and the vacuum pump will result in a lower vacuum and thus in impaired firing results For that reason the furnace head should be kept closed when the...

Страница 53: ...arious operating modes can be selected This increases the reliability of use and adjusts the furnace to the area of application The following modes are available Standard All functions of the ceramic...

Страница 54: ...he green button 4 Program screen in the Production operating mode example 2 Enter the user code Enter the user code and confirm the entry with the green button The Protected or Production modes can be...

Страница 55: ...the desired network 4 Enter the WLAN password and confirm with the green button or abort the entry with the red button 5 The WLAN connection was successfully set up After the initial set up of the WLA...

Страница 56: ...ing data must be connected 5 5 1 Image display Scroll to page 2 in the home screen and press the Images button 1 Select folder The arrow keys are used to scroll through the folder The folder can be op...

Страница 57: ...enlarge or the lower left downsize 8 Moving the image An enlarged image can be moved in all directions For that purpose press the touch screen with the finger and move the image in the desired directi...

Страница 58: ...e playback can only be stopped by pressing the corresponding button in the media player see Step 2 5 5 3 Video player Scroll to page 2 in the home screen and press the Video button 1 Selecting the fol...

Страница 59: ...sponding button is used to adjust the volume of the playback 6 Switch to full screen view The corresponding button is used to switch to the full screen view 7 Full screen view The corresponding button...

Страница 60: ...on The folder can be left by clicking the Back button The image displayed can be deleted by clicking the Delete button The analysis of the displayed image can be started by pressing the green button 1...

Страница 61: ...es the tooth to be analyzed with the reference teeth The result of the shade analysis is shown directly on the tooth to be analyzed as well as on the right side of the screen shade B1 in this example...

Страница 62: ...ull screen view By double clicking the touch screen the image section can be enlarged and reduced again 7 Verifying the designations correcting If the designations of the shade guide teeth cannot be c...

Страница 63: ...can be saved as a log file by tapping the Disk button The log file can be printed via the PrograBase X10 software or stored as a PDF file 10 Log function The current shade analysis can be saved as a l...

Страница 64: ...h on Switch on the furnace with the mains switch at the rear of the furnace The furnace conducts an automatic self test Wait until the furnace has reached the set stand by temperature and or the Optic...

Страница 65: ...urve display The IRT key see Chapter 5 1 2 is used to switch to the view of the infrared camera during the closing and predrying process The corresponding key is pressed again to return to the firing...

Страница 66: ...self test Wait until the furnace has reached the set stand by temperature Step 2 Select the program After the self test the display shows the home screen Press the Individual Programs button Step 3 En...

Страница 67: ...logy active position the objects in such a way that the infrared camera has a wide view of the objects Use the markings on the frame plate see Figure for orientation purposes The IRT key see Chapter 5...

Страница 68: ...metal frameworks with a metal margin or a holding pin have to be placed in such a way that the metal margin or the holding pin is not caught by the infrared camera as this may affect the correct tempe...

Страница 69: ...furnace head and furnace base are clean and undamaged Sealing rims of the furnace head and the furnace base weekly Check the membrane sealed keypad for visible damage If the key pad is damaged it has...

Страница 70: ...Vivadent After Sales Service if necessary 7 4 Stand by We recommend keeping the furnace head closed especially if the temperature drops below 150 C If the furnace head is open there is a risk that the...

Страница 71: ...14 Temperature in the firing chamber 410 C at the start of the calibration program i e too high Temperature too high for calibration Furnace is cooling down Try restarting the program later 16 T2 T1 E...

Страница 72: ...ion Sample might be incorrectly inserted Try again with a new sample and make sure the sample shows the correct contact 1501 yes Hint working hours The heating muffle has working hours Please perform...

Страница 73: ...used The height of the firing tray must correspond with the height of the firing trays recommended for the respective materials by Ivoclar Vivadent AG 2761 yes Error IRT Check the orientation and posi...

Страница 74: ...tart dehumidification program Incorrect or illogical temperature indication Is the thermocouple bent or fractured Contact your local Ivoclar Vivadent Service Centre Is the thermocouple correctly conne...

Страница 75: ...resetting to factory settings In this process all programs melodies screens images and volume adjustments are irrevocably reset to the factory settings For that purpose proceed as follows 1 Open Sett...

Страница 76: ...ettings The following messages are displayed Reset to factory settings successful Reset to factory settings failed To return to the home screen press either the touch button Home in the navigation bar...

Страница 77: ...le voltage fluctuations 10 Max power consumption 14 A at 100 V 12 A at 110 120 V 8 5 A at 200 240 V Acceptable data for pumps of other manufacturers Max output Final vacuum 250 W max leakage current 0...

Страница 78: ...ions Acceptable ambient temperature range 5 C to 40 C Acceptable humidity range Maximum relative humidity 80 for temperatures up to 31 C gradually decreasing to 50 relative humidity at 40 C condensati...

Страница 79: ...Centre Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the internet as PDF files Please make sure...