50
2. Start the calibration.
The display shows the last calibration value and the date of the
last calibration.
Press the START key on the membrane-sealed keypad to start the
calibration. Follow the instructions on the display.
3. Remove the firing plate.
Remove the firing plate from the furnace using the furnace tongs
and place it on the cooling tray.
Firing hours
Number of firing hours
Vacuum hours
Number of operating hours of the vacuum pump
IP address
IP address indication
Internet connection
Indicates if the furnace is connected to the Internet
MAC address
Indication of the MAC address
Press procedures
Number of all press cycles
Press cycles since last calibration
Number of press cycles since the last calibration
Video tutorials
Display of the QR code to access the video tutorials
5.5.3 Temperature calibration
The thermocouple and heating muffle of the furnace may be subject
to changes which affect the furnace temperature, depending on the
mode of operation and the frequency of use. Conduct an automatic
temperature calibration at least every six months.
Temperature calibration requires only few steps:
1. Open the temperature calibration.
Scroll to page 2 in the home screen and press the
[Temperature
Calibration]
button.
The furnace must be heated up and have reached its
stand-by temperature (403 °C) before the calibration is
started.
5. Operation and Configuration