5. Operation and Configuration
5.1.4 Explanation of the numeric pad and selection list
– Numeric pad
The numeric pad enables the entering and changing of parameters, e.g. in firing
programs or setting menus. Additionally, the currently set value as well as the
minimum and maximum values are indicated.
An entry must be confirmed with the green button. As soon as the entry has
been confirmed, the numeric pad is closed. If the button appears pale green,
the value is not in the acceptable range.
The numeric pad can be closed with the red button, without any parameters
being changed.
– Selection list
In the selection list, the desired parameter can be selected by using the up / down
arrows. An entry must be confirmed with the green button. After that, the
selection list is closed.
The selection list can be closed with the red button, without any parameters
being changed.
5.1.5 Selecting the operating mode
The furnace can be operated in two different modes:
– Pressing:
Only press programs can be selected. The stand-by temperature of the furnace is set to 700 °C.
– Firing:
Only firing programs can be selected. For Ivoclar Vivadent programs, the stand-by temperature of the furnace is set to
403 °C.
In order to switch between the press and the firing modes, you must press the CHANGE OPERATING MODE key to change
the operating mode (Chapter 5.1.2).
Once the CHANGE OPERATING MODE key has been pressed, the operating mode can be selected.
1. Select the desired operating mode.
2. After the operating mode is selected, a note is displayed indicating
the different stand-by temperatures of the respective operating modes.
Switching from “Firing” to “Pressing”:
Make sure that the furnace
has heated up to 700 °C before starting the press program.
Switching from “Pressing” to “Firing”:
Make sure that the furnace
has cooled down to 403 °C before starting the firing program.