20
IQ Power Fantom
5201292 Rev A
PROBLEM
CAUSE
SOLUTION
Fantom not working;
no indicators, fan not
running
Power not on
Turn on power switch near line
cord
Line voltage not supplied
Check line voltage and connections
Blown fuse
Check fuse and replace with
proper value / type
Warning indicator
is lit
Unit approaching control limits
Clean ion emitters with the
built-in point cleaner
Remove the ionizing module and
thoroughly clean or replace the
ionizing module
Fault indicator is lit
Unit exceeding control limits
Clean ion emitters with the
built-in point cleaner
Remove ionizing module and
thoroughly clean or replace
ionizing module
Fantom connected to Control
Station and Fantom AC power
is off
Apply line voltage to Fantom and
turn Fantom power switch on
Failure of high voltage power
supply or control circuitry
Return unit to factory authorized
repair facility
Communications
failure
Incorrectly wired cable (check
against cable illustrations in
Installation section)
Replace or repair cable
Connector problem
Check connectors for damage and
proper seating