21
5201335 REV A
IQ POWER BPS-C POWER SUPPLY
4. Remove resistant coatings deposited on static bars by wiping with isopropyl
alcohol or mineral spirits applied to a clean cloth. Apply isopropyl alcohol or
mineral spirits to a stiff nylon bristle brush and thoroughly scrub the ionization
emitter channels of the bar.
5. Blow static bars dry with clean, dry compressed air and ensure the bars are
completely dry before re-applying power to the bar.
NOTE
– Do not soak static bar or related components in alcohol or mineral
spirits. Do not use harsh solvents such as lacquer thinner, naphtha or
acetone.
WARNING – Fire Hazard
Do not turn on power supply with any trace of alcohol or mineral spirits on
the equipment. Allow all alcohol or mineral spirits to evaporate.
AVERTISSEMENT – Risque d’incendie
Ne pas installer ou d’utiliser le l’alimentation électrique à distance à
proximité de liquides ou de solvants inflammables.
Calibration
NOTE
– Calibration should be performed when the system is first installed
and may be performed after the static bar has been cleaned and the system
verified as operating correctly.
If the system is new, perform an initial calibration. The initial calibration sets relative
nominal ion output for the system. The calibration should only be performed on
systems that are new or just cleaned and known to be in proper working order.
During calibration the Conveyostat must be clear of material, i.e., no material is to
be flowing through the Conveyostat. If material is flowing in the Conveyostat (e.g.,
the machine is in operation) the calibration may be faulty.
The system should be “on” and in the operating mode (not in start-up mode). On
units NOT connected to a Control Station: press the face label on power supply
firmly on the word “Calibrate”. This will initiate the calibration sequence and set
relative nominal ion output for the system. On units connected to a Control station,
tap on the device icon, neutralizer tab. The Device Calibrate button may be found
on page 2. Tapping on Device Calibration initiates the calibration sequence.
During calibration the system output will be cycled. At completion of calibration
the indicator lights will flicker. The indicated ion output will be high. The
calibration sequence takes less than one minute.
The calibration data is stored in non-volatile memory and used on subsequent
power-ups.