About Your eXact

®

 Micro 8 Instrument

About The Accuracy/Calibration Of The Micro 8 Photometer System

The eXact

®

 Micro 8 Advanced Photometric System has been factory calibrated for your convenience. 

Customer calibration for your own special application can be performed in the Transmission Menu. You 
can expect the fixed calibrations or algorithms in the meter to be valid for the life of the meter because of 
the quality, Long-Life LED, the photo cell, and the software as written into the meter. A blue dye 
calibration solution is available for verification of meter performance.

To save power, the meter is designed to turn off after 3 minutes (timed from the last button pressed).
Should the meter turn off, the last stored zero in the meter will remain valid when the meter is turned on 
again. The eXact

®

 Micro 8 Photometer is controlled by three buttons:

1.  

ZERO/ON

: When first pressed, this button turns the meter on. When the meter is on and this button is 

pressed, it zeroes the sample in the cell. Once the meter is zeroed, this zero value applies to all menu 
parameters and is stored and retained even when meter turns off. It is recommended that each new water 
sample analyzed is zeroed before testing, to maximize accuracy.
2.  

MENU

: With each press, when the meter is on, the MENU button advances through the tests in the 

following sequence: CN1, FE2, NH3, PO4, PH5, SS6, TR7. Each test menu can store up to 20 results. To 

retrieve the stored results

, go to the desired test using the MENU key. When the desired test is displayed, 

release button and allow meter to display the last test result in that MENU. Then, 

press and hold down the 

MENU key

. Continue holding down the MENU key to scroll the stored results for that test, starting with the 

most recent result. The meter will retrieve from stored memory and display the last 20 readings in sequence 
beginning with -20, which is the latest result, followed by -19, which is the 2

nd

 latest result … finally -01, which 

is the oldest result retained. Only the last 20 readings are stored in each menu. This meter is able to store 
140 results in memory (20 in each menu).
3.  

READ

: When pressed once, this button starts a 

20 or 30 SECOND

 countdown timer depending on the 

parameter being tested. If READ is pressed a second time during the countdown, the meter exits the current 
timing and performs the next function.

If the parameter being measured is below or above the detection range, the display will show “

LO

” (Under

Range) or “

HI

” (Over Range), respectively. This feature is menu specific.

Содержание Exact Micro 8

Страница 1: ...NMENTAL EDUCATIONAL TESTING Revision 08 19 10 Part No 486800 K 2010 Industrial T est Systems Inc EXACT is a registered trademark of Industrial T est Systems Inc Rock Hill SC USA Micro 8 Advanced Photo...

Страница 2: ...ned to offer the user a more Green and cost effective alternative to testing Instead of using a 10ml water sample eXact Micro 8 uses a 4ml water sample which uses up to 60 less chemical per test The r...

Страница 3: ...new water sample analyzed is zeroed before testing to maximize accuracy 2 MENU With each press when the meter is on the MENU button advances through the tests in the following sequence CN1 FE2 NH3 PO4...

Страница 4: ...ip angled slightly move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the display disappears The meter will automatically start to count up to 600 seconds At 600...

Страница 5: ...TE If your result is 2 2 ppm or greater repeat Step 6 by using a second TPTZ Strip within the next 30 seconds This assures that all iron has reacted with adequate TPTZ reagent and that the full range...

Страница 6: ...calibration of the meter is based on a water temperature between 12 C and 31 C If temperature is below 12 C your final Phosphate value may read low FILL METER WITH SAMPLE Phosphate PO4 Test Procedure...

Страница 7: ...he calibration of the meter is based on a typical well water temperature of greater than 20 C 68 F and uses a 20 second dip see step 5 FILL METER WITH SAMPLE Sulfide as S2 Test Procedure 7 S2 SS6 1 Re...

Страница 8: ...on the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 2 Press and re press the MENU button until the display shows the parameter TR7 SELE...

Страница 9: ...ature before testing Normally the mixing cap is used for tests like Iron A stirrer such as a clean plastic coffee stirrer may also be used in place of the mixing cap Use the stirrer with gentle back a...

Страница 10: ...eXact Photometer 2 Year Limited Warranty Registration of your eXact photometer must be received within 30 days from date of purchase to activate the warranty The eXact photometer is warranted to be fr...

Страница 11: ...00 7 5 ppm 0 00 2 00 ppm 0 00 4 0 ppm 5 0 10 0 0 01 1 1 ppm Coming Soon 0 600 ppm 1 100 ppm 20 900 ppm 3 300 ppm 5 75 ppm 0 2 20 ppm 5 3 2 7 1 10 ppm Coming Soon 0 200 ppm 3 580 NTU PARAMETER No PART...

Страница 12: ...0 DR890 ppm Micro 8 ppm Total Iron results are compared using the eXact Strip Micro FE TPTZ with the eXact Reagent EZ OPEN REDUCER in the eXact Micro 8 Meter using Menu FE2 versus Hach DR890 Colorime...

Отзывы: