Cyanide Test Procedure

Remove one (1) 

eXact

®

 Strip Micro CN-1, Part No.

486812-A

 from the bottle before beginning the test.

Set the strip in a dry, convenient place and recap
the bottle immediately. Next, remove one (1) 

eXact

®

Strip Micro CN-2, Part No. 486812-B

 from the bottle

before beginning the test. Set this strip in a dry,
convenient place separate from the CN-1 strip. Recap
the bottle immediately.

REMOVE STRIP

1

6b

6a

Dip the 

eXact

® 

Strip Micro CN-1, Part No. 486812-A

into the 

CELL

 and  immediately press 

READ

. This

starts the 

30 SECOND

 countdown timer. During this

time move the strip in a gentle back and forth motion.
Because the strip is 8mm wide, the strip will need to be
angled in order to fit in the cell. Be sure test pad is fully
submerged. 

Remove and discard the strip after “1”

on the display disappears

. The cursor will move across

the display, informing you to get ready with the CN-2 strip.
When the 

30 SECOND 

countdown starts, dip

immediately the 

eXact

®

 Strip Micro CN-2, Part No.

486812-B

 into the CELL. During this time, with the strip

angled slightly, move the strip in a gentle back and forth
motion. 

Remove and discard the strip after “1” on

the display disappears

. The meter will automatically

start to count up to 600 seconds. At 600 seconds, the
cursor will move across the display while the meter
prepares to measure the sample. Record result displayed
(this result is automatically stored in CN1).

DIP STRIP AND PRESS “READ”

6

5

Press the 

ZERO/ON

 button. The cursor will move

across the display followed by 

0.00 PPM

. Sample is

ready for testing.

ZERO METER

5

4

Rinse the 

CELL

 at least 3 times with the water sample

you will be testing - rinsing minimizes the potential for
cross-contamination from a previous test. Finally, fill
cell to capacity (4ml) with the water sample. NOTE:
The calibration of the meter is based on a water
temperature between 18°C and 25°C. If temperature
is below 18°C, your final Cyanide value may read low.

FILL METER WITH SAMPLE

4

3

Press and re-press the 

MENU

 button until the

display shows the parameter 

CN1

.

SELECT TEST: CN1

3

Cn

CN1

2

Press the 

ZERO/ON

 button to power the meter on;

the display will show all annunciators,
then the current MENU selection,
followed by the last reading.

TURN METER ON

2

annunciators

(Cyanide Kit 486812 - CN-1 and CN-2 Strip)

Содержание Exact Micro 8

Страница 1: ...NMENTAL EDUCATIONAL TESTING Revision 08 19 10 Part No 486800 K 2010 Industrial T est Systems Inc EXACT is a registered trademark of Industrial T est Systems Inc Rock Hill SC USA Micro 8 Advanced Photo...

Страница 2: ...ned to offer the user a more Green and cost effective alternative to testing Instead of using a 10ml water sample eXact Micro 8 uses a 4ml water sample which uses up to 60 less chemical per test The r...

Страница 3: ...new water sample analyzed is zeroed before testing to maximize accuracy 2 MENU With each press when the meter is on the MENU button advances through the tests in the following sequence CN1 FE2 NH3 PO4...

Страница 4: ...ip angled slightly move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the display disappears The meter will automatically start to count up to 600 seconds At 600...

Страница 5: ...TE If your result is 2 2 ppm or greater repeat Step 6 by using a second TPTZ Strip within the next 30 seconds This assures that all iron has reacted with adequate TPTZ reagent and that the full range...

Страница 6: ...calibration of the meter is based on a water temperature between 12 C and 31 C If temperature is below 12 C your final Phosphate value may read low FILL METER WITH SAMPLE Phosphate PO4 Test Procedure...

Страница 7: ...he calibration of the meter is based on a typical well water temperature of greater than 20 C 68 F and uses a 20 second dip see step 5 FILL METER WITH SAMPLE Sulfide as S2 Test Procedure 7 S2 SS6 1 Re...

Страница 8: ...on the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 2 Press and re press the MENU button until the display shows the parameter TR7 SELE...

Страница 9: ...ature before testing Normally the mixing cap is used for tests like Iron A stirrer such as a clean plastic coffee stirrer may also be used in place of the mixing cap Use the stirrer with gentle back a...

Страница 10: ...eXact Photometer 2 Year Limited Warranty Registration of your eXact photometer must be received within 30 days from date of purchase to activate the warranty The eXact photometer is warranted to be fr...

Страница 11: ...00 7 5 ppm 0 00 2 00 ppm 0 00 4 0 ppm 5 0 10 0 0 01 1 1 ppm Coming Soon 0 600 ppm 1 100 ppm 20 900 ppm 3 300 ppm 5 75 ppm 0 2 20 ppm 5 3 2 7 1 10 ppm Coming Soon 0 200 ppm 3 580 NTU PARAMETER No PART...

Страница 12: ...0 DR890 ppm Micro 8 ppm Total Iron results are compared using the eXact Strip Micro FE TPTZ with the eXact Reagent EZ OPEN REDUCER in the eXact Micro 8 Meter using Menu FE2 versus Hach DR890 Colorime...

Отзывы: