Dip the 

eXact

® 

Strip Micro S

2-

, Part No. 486818-B 

into the 

CELL

 and immediately press 

READ

.

This starts the 

20 SECOND

 countdown timer. During this time move the strip in a gentle back

and forth motion. If water temperature is below 20°C, 

remove and discard the strip after “1”

on the display disappears

. If water temperature is above 20°C, 

remove and discard the strip

after “10” on the display appears

. The meter will automatically, at time “0”, start to count up

for 180 seconds. After the 180 seconds, the cursor will move across the display while the
meter prepares to measure the sample. Record result displayed (this result is automatically
stored in SS6). NOTE: Maximum detection is 1.10 ppm, however best accuracy at less than
20°C is for levels less than 0.70 ppm.

DIP STRIP AND PRESS “READ”

4

Add 4 drops of 

eXact

®

 Reagent S

2-

, Part No. 486818-A

. Press the 

ZERO/ON

 button. The

cursor will move across the display, followed by 

0.00 PPM. 

Sample is ready for testing.

ZERO METER

1

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators,

then the current MENU selection, followed by the last reading.

TURN METER ON

2

Press and re-press the 

MENU

 button until the display shows the parameter SS6.

SELECT TEST: SS6

3

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing minimizes the

potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with the
water sample. Tilt meter to discard about 0.2mL water in order to leave room for liquid reagent.
NOTE: The calibration of the meter is based on a typical well water temperature of greater
than 20°C (68°F) and uses a 20-second dip (see step 5).

FILL METER WITH SAMPLE

Sulfide (as S

2-

) Test Procedure

7

S

2-

SS6

1

Remove one (1) 

eXact

®

 Strip Micro

, part number is dependent upon the test being run,

from the bottle before beginning the test. Set the strip in a dry, convenient place and
recap the bottle immediately.

REMOVE STRIP

6

Dip the 

eXact

® 

Strip Micro (or add Reagent) 

into the 

CELL

 immediately press 

READ

. This starts

the 

20 SECOND

 countdown timer. During this time move the strip in a gentle back and forth motion.

Remove and discard the strip after “1” on the display disappears

. The cursor will move across

the display while the meter prepares to measure the sample. Record result displayed (this result is
automatically stored in TR7). After testing is completed, rinse cell immediately.

DIP STRIP AND PRESS “READ”

5

Press the 

ZERO/ON

 button. The cursor will move across the display, followed by 

100 %T

.

The sample is ready for testing.

ZERO METER

2

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators,

then the current MENU selection, followed by the last reading.

TURN METER ON

3

Press and re-press the 

MENU

 button until the display shows the parameter 

TR7

.

Some of the tests listed on the chart on page 11 will require the TR7 MENU and a
conversion chart available at www.sensafe.com.

SELECT TEST: TR7

4

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing

minimizes the potential for cross-contamination from a previous test. Finally, fill cell to
capacity (4ml) with the water sample.

FILL METER WITH SAMPLE

Transmission Test Procedure

TR

TR7

(Sulfide Kit 486818 - Reagent and Strips)

Содержание Exact Micro 8

Страница 1: ...NMENTAL EDUCATIONAL TESTING Revision 08 19 10 Part No 486800 K 2010 Industrial T est Systems Inc EXACT is a registered trademark of Industrial T est Systems Inc Rock Hill SC USA Micro 8 Advanced Photo...

Страница 2: ...ned to offer the user a more Green and cost effective alternative to testing Instead of using a 10ml water sample eXact Micro 8 uses a 4ml water sample which uses up to 60 less chemical per test The r...

Страница 3: ...new water sample analyzed is zeroed before testing to maximize accuracy 2 MENU With each press when the meter is on the MENU button advances through the tests in the following sequence CN1 FE2 NH3 PO4...

Страница 4: ...ip angled slightly move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the display disappears The meter will automatically start to count up to 600 seconds At 600...

Страница 5: ...TE If your result is 2 2 ppm or greater repeat Step 6 by using a second TPTZ Strip within the next 30 seconds This assures that all iron has reacted with adequate TPTZ reagent and that the full range...

Страница 6: ...calibration of the meter is based on a water temperature between 12 C and 31 C If temperature is below 12 C your final Phosphate value may read low FILL METER WITH SAMPLE Phosphate PO4 Test Procedure...

Страница 7: ...he calibration of the meter is based on a typical well water temperature of greater than 20 C 68 F and uses a 20 second dip see step 5 FILL METER WITH SAMPLE Sulfide as S2 Test Procedure 7 S2 SS6 1 Re...

Страница 8: ...on the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 2 Press and re press the MENU button until the display shows the parameter TR7 SELE...

Страница 9: ...ature before testing Normally the mixing cap is used for tests like Iron A stirrer such as a clean plastic coffee stirrer may also be used in place of the mixing cap Use the stirrer with gentle back a...

Страница 10: ...eXact Photometer 2 Year Limited Warranty Registration of your eXact photometer must be received within 30 days from date of purchase to activate the warranty The eXact photometer is warranted to be fr...

Страница 11: ...00 7 5 ppm 0 00 2 00 ppm 0 00 4 0 ppm 5 0 10 0 0 01 1 1 ppm Coming Soon 0 600 ppm 1 100 ppm 20 900 ppm 3 300 ppm 5 75 ppm 0 2 20 ppm 5 3 2 7 1 10 ppm Coming Soon 0 200 ppm 3 580 NTU PARAMETER No PART...

Страница 12: ...0 DR890 ppm Micro 8 ppm Total Iron results are compared using the eXact Strip Micro FE TPTZ with the eXact Reagent EZ OPEN REDUCER in the eXact Micro 8 Meter using Menu FE2 versus Hach DR890 Colorime...

Отзывы: