5
Dip the
eXact
®
Strip Micro PO
4
, Part No. 486814
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle
back and forth motion.
Remove and discard the strip after “1” on the display disappears
.
The meter will automatically start to count down for 120 seconds. After the 120 seconds, the
cursor will move across the display while the meter prepares to measure the sample. Record
result displayed (this result is automatically stored in PO4).
DIP STRIP -
(read carefully and follow procedure closely)
4
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
Sample is ready for testing.
ZERO METER
1
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the
MENU
button until the display shows the parameter PO4.
SELECT TEST: PO4
3
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with
the water sample. NOTE: The calibration of the meter is based on a water temperature between
12°C and 31°C. If temperature is below 12°C, your final Phosphate value may read low.
FILL METER WITH SAMPLE
Phosphate (PO
4
) Test Procedure
6
PO
4
PO4
BT-pH Test Procedure
BT-PH
PH5
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Dip the
eXact
®
Strip Micro BT-PH, Part No. 486652
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion.
Remove and discard the strip after “1” on the display disappears
. The
cursor will move across the display while the meter prepares to measure the sample. Record
result displayed (this result is automatically stored in PH5).
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display followed by
0.00
.
Sample is ready for testing.
ZERO METER
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the
MENU
button until the display shows the parameter
PH5
.
SELECT TEST: PH5
2
(Phosphate Kit 486814 - Strips only)
(BT-pH Kit 486652 - Strips only)