The setting principles of calculation parameters in other modes are the same.
Formula parameter descriptions:
Name
Description
V
in1
Indicates the start voltage input to pin 8. The
setting range is from –10 to 10.
V
in2
Indicates the end voltage input to pin 8. The
setting range is from –10 to 10, and V
in2
>
V
in1
.
V
out1
The starting value of the input voltage in CV
mode.
V
out2
The end value of the input voltage in CV
mode, and V
out2
> V
out1
.
Analog Control
The pin connection method and instrument operation method in four basic
modes of the load are the same. The voltage control in CV mode is taken as an
example below to describe the connection and usage.
1. Refer to the figure below to complete the pin connection.