WatchPAT™200 System
69
Operation Manual
Table 4 - from IEC 60601-1-2:2007
Guidance and manufacturer's declaration
– electromagnetic immunity – WP200
The
WP 200
is intended for use in the electromagnetic environment specified below; The customer or the user of the
WP200
should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 k Hz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
WP200
, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter
.
Recommended separation distance
d
= 1.17
√
P
d
= 1.17
√
P 80 M Hz t o 800 MHz
d=
2.3
√
P 800 MHz t o 2,5 GHz
where
P
i s the maximum output power rating of the
transmitter in watts (W)according to the transmitter
manufacturer and
d
i s the recommended separation
Distance in meters (m).
Field strengths from fixed R F transmitters, as
determined by an electromagnetic site survey ,
a
should be
less than the compliance level in each frequency
range .
d
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations .Electromagnetic propagation is affected by absorption
And reflection from structures objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the
WP200
is used exceeds the applicable RF compliance level above, the
WP200
should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the
WP200
.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.