26
1
2
180°
Lamp replacement
The replacement of the lamps must be carried out according to the working hours due to the decrease in efficiency of the
same. The change is required every 9000 hours of work which correspond to approximately one year of continuous service.
Before starting this type of maintenance, make sure that the equipment is disconnected from the power supply.
To access the lamps it is therefore sufficient to remove the protective gray fixed by means of 4 blind nuts
once the lamp is accessed, it can be easily replaced by rotating it 180 ° before extracting it from the support feet, as in the
case of a common neon tube.
Subsequently, the new lamps are inserted by acting in reverse.
In order to maintain the sanitizing and germicidal efficiency of the Stril Room components, it is essential to strictly follow
the maintenance schedule indicated below.
CAUTION!! Before any type of maintenance, especially internal ones, it is mandatory to disconnect the
components from the power supply network.
For some of the maintenance operations it is necessary to refer to the working hours of the lamps which are
stored by means of the hour meter located inside the control console.
Recommended maintenance schedule
Operation description
working time / hours
Cleaning the reflective support
At least once a month in order to keep the reflective walls
clean for maximum efficiency
Change lamps + lamp holder
every 9000 hours (approximately one year in continuous
service)
Replacement of the IR sensor
every 9000 hours and in any case depending on the state of
wear of the external lens which could compromise correct
presence detection
MAINTENANCE