48/54
IWQ Series Service Manual
Chapter 5. Maintenance
5.1 Periodical Replacement Parts
•
No periodical parts replacement is needed against five-year product life.
•
By disassembling the rear cover for maintenance service, the airtight quality of the gasket packing
decreases. Replace with a new gasket packing.
Gasket seal washer
Rear cover
Gasket packing
Front case
Sealing
screw
Seal
5.2 Replacing and Adjusting Load Cell
Specifications
Type
Model code
Rated output
Input
resistance
Output
resistance
Insulation
resistance
IWQ-150
LC'AT-200'
2.3mV/V
±
15%
2000
∼
2600
Ω
2000
∼
2600
Ω
10G
Ω
IWQ-30
LC'AT-50'
2.3mV/V
±
15%
2000
∼
2600
Ω
2000
∼
2600
Ω
10G
Ω