36/54
IWQ Series Service Manual
Address Weight
Data Contents
Data Range
112A
-
Starting range
0: ±1/50 of weighing capacity (±2%)
1: ±1/25 of weighing capacity (±4%)
2: ±1/10 of weighing capacity (±10%)
3: ±1/7.5 of weighing capacity (±13.3%)
4: ±1/6 of weighing capacity (±16.6%)
5: ±1/5 of weighing capacity (±20%)
6: ±1/4 of weighing capacity (±25%)
7: ±1/3 of weighing capacity (±33.3%)
8: ±1/2 of weighing capacity (±50%)
9
∼
Not available.
112B
-
Stabilized,
restabilized
frequency
0
∼
15 times.
113A
-
Stabilized/re-
stabilized range
n=0
∼
15 (
±
n/10 e )
113B
-
Re-stabilization
starting range
n=0
∼
15 (
±
n/10 e )
114A
1
Zero point mark
0: Lights on at true zero, 1: Lights on at provisional zero
2
Over-scale indication 0: BLANK
2: “OL”
4
Below true zero
indication
0: “
_____
”
4: Negative value
8
Decimal point
indicator
0: “.”
8: “,”
114B
1
Over-scale range
0: Indication up to +9e 1: Indication up to +3e
2
Tare deduction
0: Possible
2: Not possible
4
Clearing tare weight
by pressing ZERO
key
0: No
4: Yes
8
Zero suppress
function (at multi-
interval spec.)
0: Available
4: Not available
115A
1
Input value/preset
tare weight deduction
and preset unit
weight function
0: Available
8: Not available
2
Zero tracking
0: Available
2: Not available
4
Micro weight follow-
up
0: Available
4: Not available
8
Unstable range
0: ±0.5e
8: ±20e
115B
1
Number of preset
(Area adjustment)
<Printer output form>
0: 10 items
1: 5 items
(available)
(not available)
<with ’PT’,’N’ prints>
<without ‘PT’,’N’ prints>
2
Counting function
0: Available
2: Not available
4
Weight unit cursor
indication (at normal
switching lb-kg)
0: Normally lighting
4: Lighting shutoff for non
switching
(Not available)
(Available)
8
Reserve