12/54
IWQ Series Service Manual
Set No.
Setup title
Set value & functions
(Default underlined)
Remarks
F11
Serial output
method
0: No output
1: Send-out in discharge
(synchronous to A/D)
2: Output conditions (F12)
3: As requested for DATA input (RS
input) (contact input)
•
Set other than 0 for linked
printer.
•
When “2: Output conditions” is
selected, the conditions
selected in F12 are applicable.
F12
Serial output
conditions
0: Output when automatic/within range
/re-stabilize
1: Output when manual/within
range/re-stabilize
2: Output when automatic/range
disregarded/re-stabilize
3: Output when manual/ range
disregarded/re-stabilize
4: Output when automatic/within
range/zero return
5: Output when manual/within
range/zero return
6: Output when automatic/range
disregarded/zero return
7: Output when manual/range
disregarded/zero return
•
When “Send-out in discharge” is set,
the O8 command is not available.
•
When manual printing is set, key
input is determined to be error for a
unstable state.
•
Functions when selecting 2:
output conditions for F11.
•
”Within range” means the range
between upper and lower limit
values.
•
”Range disregarded” means
free from the range between
upper and lower limit values
even if they are set.
•
Applicable to either within or
outside the range, when upper
and lower limits are not set.
Items relating to relay output (relay box)
Set No.
Setup title
Set value & functions
(Default underlined)
Remarks
OUT1
OUT2
OUT3
OUT4
0
Stable
Zero point
marking
Accept
Over
1
Stable
Zero
display
Accept
Over
2
Stable
Zero
display
Accept
Under &
Over
3
Stable
Under
Accept
Over
F13
Setting contact
output signal
(relay box)
4
Tare
deduction
Under
Accept
Over
•
Output is ON
for up each of
the conditions.
•
Signals are all
OFF when
power is ON.
F14
Selecting
relay input
0: “ON/OFF” key
1: “TARE” key
2: “ZERO” key
3: Scale data output (same response as O8 command)