IWQ Series Service Manual
3.6.2 Measurement conditions setup table
Address Weight Data Contents
Data Range
110A
1
0: 0 1:0.0
2
Decimal point
position
2: 0.00 3:0.000
4
Minimum
indication
0: 1 (At dual range/multi-interval 1
−
2 switching)
4: 2 (At dual range/multi-interval 2
−
5
−
10 switching)
8
8: 5 (At dual range/multi-interval 5
−
10
−
20 switching)
C: Not available
110B
-
Accuracy
(resolution)
0: 1/500
1:1/600
2:1/750
3:1/1000
4: 1/1200 5:1/1500 6:1/2000
7:1/2500
8: 1/3000 9:1/4000 A:1/5000
B:1/6000
C: 1/7500 D:1/10000 E:1/12000
F:1/15000
1
2
4
Switching
method
0: Single Range, checker mode when power is turned ON.
1: 1/3000 Multi-Interval, checker mode when power is turned
ON.
2: Single Range, checker/counting mode maintained.
3: 1/3000 Multi-Interval, checker/counting mode maintained.
4: Single Range (bathroom scale specifications)
5: 1/3000 Multi-Interval (A/B range switching)
6: Single Range (lb/kg switching)
7: 1/3000 Multi-Interval (lb/kg switching)
<Except for 2 and 3, the mode not maintained by the CHG
key>
111A
8
1
Weight unit
(applicable only
to the trans-
mission
function)
<111A> <111B>
0
+
0 : kg (kg-lb)
8
+
0 : g
0
+
1 : lb
8
+
1 : oz
2
4
111B
8
Filter setup
Filter conditions can be selected to ease sampling time of
weight shift and flickering.
Optimal condition is determined by scale No.
When changed, weight may not be measured correctly.
(Cut off)
(Notch)
(Output rate)
0: 0.66 Hz
2.50 Hz
400 ms
2: 0.84 Hz
3.20 Hz
312 ms
4: 1.05 Hz
4.00 Hz
250 ms
6: 1.31 Hz
5.00 Hz
200 ms
8: 1.68 Hz
6.40 Hz
156 ms
A: 2.10 Hz
8.00 Hz
125 ms
C: 2.62 Hz
10.00 Hz
100 ms
E: 3.35 Hz
12.80 Hz
78 ms