2 Technical data
2.1 EHS Standard series
SPECIFICATIONS
EHS CG
EHS CFG
EHS FG
Polarity
Factory fixed, positive or negative
Floating principle
Common Ground
Common Floating Ground
Single Floating Ground
Potential difference
none
56 V channel/GND
20 V channel/channel/GND,
optionally up to 2 kV
Ripple and noise (f > 10 Hz)
< 10 - 20 mV
p-p
optionally VLN: < 3 - 5 mV
p-p
< 10 mV
p-p
Ripple and noise (f > 1 kHz)
< 2 − 3 mV
Stablity
Stability – [ΔV
out
vs. ΔV
in
]
< 1 • 10
-4
• V
nom
Stability – [ΔV
out
vs. ΔR
load
]
< 5 • 10
-4
• V
nom
< 2 • 10
-4
• V
nom
Long term stability
(1h warmup) 24h
< 5 • 10
-5
• V
nom
Temperature coefficient
< 50 ppm / K
Resolution voltage setting
2 • 10
-6
• V
nom
4 • 10
-5
• V
nom
Resolution current setting
2 • 10
-6
• I
nom
4 • 10
-5
• V
nom
Resolution voltage measurement
(1
2 • 10
-6
• V
nom
Resolution current measurement
(1
2 • 10
-6
• I
nom
Measurement accuracy
–
The measurement accuracy is guaranteed in the range 1% • V
nom
< V
out
< V
nom
and for 1 year
Accuracy voltage measurement
± (0.01 % • V
out
+ 0.02 % • V
nom
)
Accuracy current measurement
± (0.02 % • I
out
+ 0.02 % • I
nom
)
Sample rates ADC (SPS)
5, 10, 25, 50, 60, 100,
500
(2
5, 10, 25,
50
(2
, 60
Digital filter averages
1, 16,
64
(2
, 256, 512, 1024
Voltage ramp up / down
up to 0.2 • V
nom
/s
up to 0.2 • V
nom
/ s | optionally up to 0.75 • V
nom
/ s
Hardware limits
Potentiometer per module [V
max
and I
max
]
Limit monitor voltage
2.5 V
Digital interface
CAN
Protection
Safety loop, overload and short circuit protected,
optionally INHIBIT per channel (ID / IU, NID / NIU)
(
ATTENTION:
t
here is only one short circuit or arc per second allowed!)
HV connector
R51 | SHV
System connector
96 PIN (MMS HV compatible)
Safety loop connector
Lemo 2pole (SL),
Limit monitor connector
Lemo 1pole,
Lemo 2pole,
Lemo 1pole,
Case
19” plug-in cassette
Dimensions – L/W/H
220mm / 8HP / 6U
EHS High Voltage Power Supply series | Last changed on: 2022-03-14 |
www.iseg-hv.com
10/51