4.2 SLA – Active safety loop
Actively opens the Safety loop in case of a trip or a delayed trip. This option allows to shut down other modules and devices by
interrupting the SL when a trip is detected.
4.3 SLP – Internally powered safety loop
Internal current source for the Safety Loop (no galvanic isolation of the SL and the crate GND).
4.4 1CR – One current measurement range only (HP series)
Only one current measurement range for High Precision Modules.
4.5 F02 – High floating voltage
200 V isolation for Modules with Floating Ground (FG).
With option “F02” the floating voltage is internally not limited. The user is responsible to limit potential differences between
individual channel – GNDs and Crate – GND. Exceeding the isolation voltage can damage the module.
4.6 F20 – Very high floating voltage
2.000 V isolation for Modules with Floating Ground (FG).
With option “F20” the floating voltage is internally not limited. The user is responsible to limit potential differences between
individual channel – GNDs and Crate – GND. Exceeding the isolation voltage can damage the module.
4.7 TC – Lower temperature coefficient
Improved temperature coefficient of 10ppm/K (T10). This is only for 8 channels common floating ground modules available.
4.8 VLN
Reduced ripple see chapter 2 Technical data.
4.9 L – Lower output current (HP only)
The output current is limited to a lower value, e.g. 100 µA. With this option only one current measurement range available.