15510-EN Rev G
Maintenance 31
Power Measurements using a CW Delivery Device
Power (mW) Exposure Duration (ms)
Meter Reading (mW)
Acceptable Range (mW)
200
1000–3000
160–240
500
1000–3000
400-600
1000
1000–3000
800-1200
2000
1000–3000
1600-2400
Data for power measurement equipment: _______________
Calibration date: _______________
Meter Model and Serial Number: _______________
Calibrated by: _______________
Power Measurements using a MicroPulse
®
Delivery Device
Exposure
Duration (ms)
MicroPulse
®
Duration (ms)
MicroPulse
®
Interval (ms)
Indicated Power
(mW)
Measured
Power (mW)
Acceptable
Range (mW)
1000–3000
1.0
1.0
200
80-120
1000–3000
1.0
1.0
500
200-300
1000–3000
1.0
1.0
1000
400-600
1000–3000
1.0
1.0
2000
800-1200
Data for power measurement equipment: _______________
Calibration date: _______________
Meter Model and Serial Number: _______________
Calibrated by: _______________