15510-EN Rev G
Operation 23
Options Screen
To access the Options screen, touch
o
ptionS
.
A
Set aiming beam in Standby: ON or OFF.
B
Set aiming beam in Treat:
• OFF: Aiming beam OFF while footswitch is depressed.
• ON: ON at all times.
• Blink: Blink at fixed rate (not synchronized with laser settings).
C
Set voice prompt: Female, Male, OFF. Use only when adjusting power with footswitch.
D
Set Auxiliary: ON in Standby or ON in Treat. Operate a warning light or auditory signal outside the
treatment room.
E
Press bar to select it (white=active bar). Use control knobs to set volume.
F
Press bar to select it (white=active bar). Use control knobs to set brightness.
G
Discard changes and return to Treat screen.
H
Save changes and return to Treat screen.
E
A
B
C
D
G
H
F