15510-EN Rev G
Setup 13
Choosing a Location
Choose a well-ventilated location within the specified operating range of the console.
Place the laser system on a table or on existing operating room equipment. Allow at least 5 cm (2 in.)
of clearance on each side.
In the US, this equipment must be connected to an electrical supply source at 120V or 240V with a
center tap.
To ensure that all local electrical requirements can be met, the system is equipped with a
hospital-grade (green dot) three-wire grounding plug. When choosing the location, ensure that a
grounding-type AC outlet is available; it is required for safe operation.
The power cord included in the packaging is appropriate for your location. Always use an approved
three-wire grounding cord set. Do not alter the power inlet. To ensure proper grounding, follow local
electrical codes before installing the system.
CAUTIONS:
Do not defeat the purpose of the grounding pin. This equipment is intended to be electrically
grounded. Contact a licensed electrician if your outlet prevents you from inserting the plug.
Do not position or use the system near open flames.
Connecting the Components
CAUTION: Do not connect two footswitches to the laser console.
NOTES: Refer to the appropriate delivery device manual for specific connection instructions.
The Auxiliary Output contact supports low voltage electrical signaling circuits of up to
5 amps and 24V AC or DC. Ensure that all wiring conforms to local electrical codes.