Out of range behavior
-Remote control can’t operate on the receiver pods.
-Receiver pods won’t work.
Wireless Coexistence Testing
The PlayMakar
®
Therapeutic Wearable System™ protocol is designed to allow coexistence with at least 4
other PlayMakar
®
devices.
Each PlayMakar
®
Therapeutic Wearable System™ remote control and receiver pod is assigned a unique ID.
Before initial use, the remote is paired with 4 receiver pods in its network.
Designed for coexistence with wireless
products in the same RF band
-Bluetooth (IEEE 802.15.1)
-Wi-Fi (IEEE 802.11)
Utilizes established coexistence princi-
ples to minimize cross-talk with other
wireless devices
-FDMA (Frequency Division Multiple Access)
-TDMA (Time Division Multiple Access)
Interference from other devices
-If the PlayMakar
®
Therapeutic Wearable System™'s RF channel is
interfered by emission from other devices, fail-safe outcome is that
iReliev's remote control will not operate, and/or reset the stimulation
level to “0”; the receiver pods will stop stimulation.
-Interference from other RF wireless and mobile communication
devices is possible. Refer to Table 1 for recommended distance be-
tween iReliev's PlayMakar
®
devices with other RF devices.
CONFORMITY OF STANDARDS
• The PlayMakar
®
Therapeutic Wearable System™ complies with current medical standards.
• The PlayMakar
®
Therapeutic Wearable System™ also complies with IEC 60601-1 on general safety require-
ments for electromedical devices, the IEC 60601-1-2 standard on electromagnetic compatibility, the IEC 60601-2-
10 standard on particular safety requirements for nerve and muscle stimulators, and the IEC 60601-1-11 standard
for use in the home environment.
33
www.iReliev.com