Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d= 0.6
P
80 MHz to 800 MHz
d= 0.35
P
800 MHz to 2.5 GHz
d= 0.7
P
0.01
0.06
0.04
0.07
0.1
0.19
0.11
0.22
1
0.6
0.35
0.7
10
1.9
1.1
2.2
100
6
3.5
7
Declaration – electromagnetic emissions and immunity for EQUIPMENT and SYSTEMS that are not LIFESUP-
PORTING and are specified for use only in a shielded location
Table 2 The PlayMakar
®
Therapeutic Wearable System
™
declaration – electromagnetic immunity
The PlayMakar
®
Therapeutic Wearable System™ is intended for use in the electromagnetic environment
specified below.
The customer or the user of the PlayMakar
®
Therapeutic Wearable System™ should assure that it is used in
such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6V
Portable and mobile RF communications equipment
should be used no closer to any part of the EQUIP-
MENT or SYSTEM including cables, than the recom-
mended separation distance calculated from the
equation applicable to the frequency of the trans-
mitter.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
10V/m
35
www.iReliev.com