Power frequen-
cy (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
The magnetic field from common ap-
pliances are not expected to affect the
device.
▲
NOTE: U
T
is the a.c. main voltage prior to application of the test level.
Table 4 Declaration – electromagnetic emissions
The PlayMakar
®
Therapeutic Wearable System™ is intended for use in the electromagnetic environment
specified below. The customer or the user of the system should assure that it is used in such an environ-
ment.
Emissions test
Compliance
Electromagnetic environment guidance
CE emissions CISPR11
Group 1
The PlayMakar
®
Therapeutic Wearable System™
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RE emissions CISPR11
Class B
The PlayMakar
®
Therapeutic Wearable System™
is suitable for use in all establishments, including
domestic establishments and those directly con-
nected to the public low-voltage power supply
network that supplies buildings used for domes-
tic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions IEC 61000-3-3
Complies
37
www.iReliev.com