121
YLS-K
28.
If there is dirt on the quartz-block end-face, remove the screwed-
on protective glass and follow the cleaning steps described under
the protective conduit option.
29.
If the quartz-block end-face does not exhibit any impurities, ad-
just the microscope's focus to the protective glass surface (first fo-
cusing plane).
30.
If the protective glass is dirty, perform the cleaning similar to the
cleaning procedure described under the protective conduit op-
tion.
31.
Alternatively, replace the dirty protective glass for a clean protec-
tive glass. In doing so, ensure that no dirt is situated on the
quartz-block end-face. Do not touch the new protective glass with
your fingers. Use the white protective packaging in which the pro-
tective glass was delivered in order to screw the new protective
glass onto the fiber connector.
9.1.2 Measuring the laser output power
WARNING
Danger from laser radiation
Depending on the measurement method, there is a
danger from laser radiation during measurement of
the output power as well as the beam quality. Laser
radiation can damage the eyes and skin.
ð
Ensure that the measurement below is per-
formed only by IPG service personnel.
Test point
Laser output power
Test method
Performance measurement
Time interval
Annually
Time duration
2 hours / depending on the laser design
Tools
Performance measuring device
Performance
IPG Laser GmbH
Contact
The contact information of the competent IPG service
department can be found in the paragraph
.
| Maintenance