4-4
Figure 4-2. TF0088 Assembly Instructions
(20) Turn the Power Switch on the
Millennia
®
3155MVS/3155A
monitor to the OFF
(
F
) position.
(21) Disconnect TF088 and DMM leads from the unit.
(22) Reconnect the battery cable from AB72A to AB69A J1.
(23) Reassemble the unit as required.
4.4
Agents Adjustment Procedure (Millennia
®
3155A Only).
On the
Millennia
®
3155A
monitor, perform the following procedure to adjust the Agents System.
a.
Remove top cover and front panel assembly (Reference Section 3: Disassembly
Procedures). Remove the shield (leave cables connected).
b.
Turn the monitor on.
c.
Setup Parameter Selections as follows:
(1)
Press the
SETUP
Control Key.
(2)
Turn the Rotary Knob to highlight the
PARAMETER SELECTION
menu
selection.
(3)
Press the Rotary Knob to select
PARAMETER SELECTION
.
(4)
Turn the Rotary Knob to highlight the
ETCO2
menu selection.
(5)
Press the Rotary Knob to select
ETCO2
.
(6)
Turn the Rotary Knob to highlight the
OFF
menu selection.
(7)
Press the Rotary Knob to select
OFF
.
(8)
Press the
NORMAL SCREEN
Control Key.
d.
Enter Monitor Calibration Mode as follows:
(1)
Press the
SETUP
Control Key.
(2)
Turn the Rotary Knob to highlight the
SERVICE (BIO-MED)
menu selection.
Содержание Millennia 3155A
Страница 2: ......
Страница 4: ......
Страница 12: ...viii THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...2 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 52: ...5 10 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 54: ...6 2...
Страница 55: ...6 3...
Страница 56: ...6 4...
Страница 57: ...6 5...
Страница 58: ...6 6...
Страница 59: ...6 7...
Страница 60: ...6 8...
Страница 61: ...6 9...
Страница 62: ...6 10...
Страница 63: ...A 1 APPENDIX A LIST OF SYMBOLS...
Страница 64: ...A 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ...C 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 71: ...NOTES...
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......