Technical Data
Revised October 2004
3800 Vital Signs Monitor Service Manual
13-5
Respiration
Method
Impedance Pneumography.
Input Connector
Same as ECG.
Sweep Rate
Fixed at 6.25 mm/second.
Sensing Lead
II.
RR Resolution
1 bpm (breaths per minute).
Measurement Range
2 to 120 bpm.
Measurement Accuracy
±3 bpm.
Measurement Sensitivity
0.25 ohms (minimum).
Report Interval
1 second.
Bandwidth
0.17 to 3.3 Hz (-3dB).
Impedance Measuring Current < 10 uA @ 44 kHz square wave across Lead II.
Pulse Oximetry
Method
Absorption – Spectrophotometric (dual wavelength)
(Functional oxygen saturation of arterial hemoglobin).
Input Connector
7-pin Limo connector. See Accessories for probe information.
SpO2 / PR Resolution
SpO2: 1 O2%.
PR: 1 bpm (beat per minute).
Measurement Range
SpO2: 1 to 100%.
PR: 25 to 240 bpm.
Measurement Accuracy
SpO2: from 70 to 100%: ±2% (O2%), < 70%: unspecified.
PR: ±3 bpm (beats per minute).
Measurement Test Method
BioTek Instruments Index2 SpO2 Simulator.
Report Interval
1 second. Numeric values held < 30 seconds, per EN 865.
Pulse Tone
Yes (pulse tone pitch tied to SpO2 parameter value).
YSI Temperature
Compatibility
YSI 400-series probes.
Input Connector
2-pin connector.
Display Units
°F and °C (user-selectable).
Measurement Resolution
0.1 °F (0.1 °C).
Measurement Range
32.0 to 122.0 °F (0.0 to 50.0 °C).
Measurement Accuracy
±0.1 °F (±0.1 °C) plus probe tolerance.
Содержание Escort M8
Страница 1: ......
Страница 2: ......
Страница 26: ......
Страница 32: ......
Страница 34: ......
Страница 44: ......
Страница 55: ...Repair Revised October 2004 3800 Vital Signs Monitor Service Manual 7 11 Figure 7 10 Input Panel Assembly View...
Страница 56: ...Repair 7 12 3800 Vital Signs Monitor Service Manual Revised October 2004 Figure 7 11 Recorder Subassembly View...
Страница 60: ......
Страница 66: ......
Страница 72: ......
Страница 78: ......
Страница 80: ......