Cleaning
Revised October 2004
3800 Vital Signs Monitor Service Manual
10-3
CAUTION -
If monitor becomes accidentally wet during use, discontinue operation of the
monitor until all affected components have been cleaned and permitted to dry completely.
Contact your local InvivoMDE representative if additional informatioin is required.
CAUTION -
Always disconnect the Escort M8 Vital Signs Monitor from AC mains power
before cleaning.
CAUTION -
Do not use harsh chemicals for cleaning – in particular, do not use
disinfectants that contain phenol as they can spot plastics. Do not steam autoclave, gas sterilize,
irradiate, subject to intense vacuum, or immerse in water or cleaning solution. Be careful to
avoid getting cleaning liquids into connectors or the monitor. If this occurs, allow the unit to
dry in warm air for 2 hours, then check to make sure all monitoring functions are still working
properly.
CAUTION -
Take particular care when cleaning the NIBP cuff, NIBP hose, and NIBP
connector on the Escort M8 Vital Signs Monitor to prevent fluid from entering the connectors.
Fluid in the NIBP airway may affect blood pressure determination accuracy and damage the
monitor.
CAUTION -
Accessories that fall on the floor should be inspected for contamination and
proper functionality. If contamination is observed, then this cleaning procedure should be
followed.
CAUTION -
The user has the responsibility to validate any deviations from the
recommended method of cleaning and disinfection.
CAUTION -
Do not reuse single use items.
Содержание Escort M8
Страница 1: ......
Страница 2: ......
Страница 26: ......
Страница 32: ......
Страница 34: ......
Страница 44: ......
Страница 55: ...Repair Revised October 2004 3800 Vital Signs Monitor Service Manual 7 11 Figure 7 10 Input Panel Assembly View...
Страница 56: ...Repair 7 12 3800 Vital Signs Monitor Service Manual Revised October 2004 Figure 7 11 Recorder Subassembly View...
Страница 60: ......
Страница 66: ......
Страница 72: ......
Страница 78: ......
Страница 80: ......