Configuring the Monitor
Revised October 2004
3800 Vital Signs Monitor Service Manual
3-3
Figure 3-5. Setup Alarm System Menu
3.
Setup Alarm System Menu.
To access the Alarm System Setup Menu, rotate the
Rotary Knob until the Alarms option is highlighted then press the knob. This menu
provides the operator with options to select the Alarm Silence Time, select the
Alarm Suspend Time, turn Alarm Validation On or Off, turn HR Alarms Off
Message On or Off, turn the SpO2 Alarms Off Message On or Off and set the
Second Speaker Time.
Setting Name
Default Value
Possible Values
Alarm Silence Time
1 minute
1 - 4 minutes
Alarm Suspend Time
1 minute
1 - 4 minutes
Alarm Validation
On
Off, On
HR Alarms Off Message
On
Off, On
SpO2 Alarms Off Message
On
Off, On
Second Speaker Time
2 minutes
0 - 3 minutes