Prepare the gel
1.
Remove the gel from the package.
2.
Gently remove the combs. Do not allow the combs to bend or create suction in the wells during
removal.
3.
Place the cassette into a E
‑
Gel
™
Adapter.
4.
Insert the gel cassette and adapter into the E
‑
Gel
™
Power Snap Plus Electrophoresis Device,
starting from the right edge.
5.
Press down on the left side of the cassette to secure it into the device.
6.
Load gels within 15 minutes after opening the package.
Sample loading guidelines
•
Use the recommended total loading volume for each gel type.
Do not load more than
recommended amount of DNA sample or ladder per well.
•
Load deionized water into all empty wells.
•
Keep all sample volumes uniform.
If you do not have enough samples to load all the wells of
the gel, load an identical volume of deionized water into any empty wells. Prepare your samples
by adding E
‑
Gel
™
1X Sample Loading Buffer or deionized water to the required amount of DNA to
bring the total required sample volume.
• Avoid introducing bubbles while loading. Bubbles can cause band distortion.
Gel type
Total loading volume
E
‑
Gel
™
EX
20 µL
E
‑
Gel
™
EX Double Comb
E
‑
Gel
™
with SYBR
™
Safe
E
‑
Gel
™
Double Comb with SYBR
™
Safe
E
‑
Gel
™
48 Agarose Gel with SYBR
™
Safe
15 µL
E
‑
Gel
™
96 Agarose Gel with SYBR
™
Safe
20 µL
E
‑
Gel
™
CloneWell
™
II
25 µL
E
‑
Gel
™
SizeSelect
™
II
25 µL
E
‑
Gel
™
NGS
20 µL
E-PAGE
™
48
15 µL
E-PAGE
™
96
20 µL
Chapter 2
Methods
Using the E
‑
Gel
™
Power Snap Plus Electrophoresis Device
2
E
‑
Gel
™
Power Snap Plus Electrophoresis System User Guide
27