TYPICAL PRODUCT PARAMETERS
Part No 1164934
9
Perfecto
2
™
V
*NOTE:
Stated
concentration
levels
achieved
after
initial
warm
‐
up
period
(approximately
30
minutes).
IRC5PO2V INDICATORS
Regulatory Listing:
Double Insulated Product ETL certified complying with UL 1097
ETL certified to CSA C22.2 No. 68.
Electrical:
No extension cords.
Placement:
No closer than three inches from any wall, furniture, draperies,
or similar surfaces.
Tubing:
7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO
NOT pinch).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Recommended Storage and Shipping Temper-
ature:
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
LABEL
SYMBOL
O
2
PURITY
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
O
2