SECTION 6—PREVENTIVE MAINTENANCE
Part No 1164934
29
Perfecto
2
™
V
Cleaning the Heat Exchanger
DANGER
To
prevent
electrical
shock,
ALWAYS
disconnect
concentrator
from
electrical
outlet
before
servicing.
NOTE:
For
this
procedure,
refer
to
FIGURE 6.6.
1. Unplug
unit.
2. Remove
cabinet.
Refer
to
Removing
Cabinet
on
page 23.
CAUTION
Use
care
not
to
deform
heat
exchanger
when
installing,
removing
or
cleaning.
3. Remove
excess
dirt
using
compressed
air
or
vacuum.
4. Reinstall
cabinet.
Refer
to
Cabinet
on
page 23.
FIGURE 6.6
Cleaning the Heat Exchanger
Heat Exchanger