background image

SECTION 2—INSTALLATION/SEQUENCE OF OPERATION

Part No 1164934

13

Perfecto

2

V

The

 

electrical

 

activation

 

of

 

the

 

4

way

 

Valve

 

is

 

accomplished

 

every

 

8

 

to

 

15

 

seconds

 

by

 

the

 

pressure

 

sensor

 

and

 

P.C.

 

Board

 

electronics

 

when

 

the

 

pressure

 

reaches

 

a

 

set

 

point

 

of

 

21

 

p.s.i.

 

(144.79

 

kPa)

 

output

 

flows

 

4

 

L/min

 

and

 

above

 

or

 

16

 

p.s.i.

 

(110.32

 

kPa)

 

output

 

flows

 

3

 

L/min

 

and

 

below.

 

The

 

time

 

between

 

cycles

 

is

 

dependent

 

on

 

altitude,

 

flow

 

rate

 

and

 

internal

 

environmental

 

factors.

A

 

P.E.

 

valve

 

opens

 

just

 

prior

 

to

 

the

 

shift

 

of

 

the

 

4

way

 

valve.

 

This

 

allows

 

highly

 

concentrated

 

oxygen

 

to

 

enter

 

the

 

just

 

exhausted

 

bed

 

from

 

the

 

top.

 

This

 

additional

 

pressure

 

allows

 

the

 

bed

 

to

 

start

 

its

 

cycle

 

at

 

a

 

higher

 

pressure.

 

The

 

P.E.

 

valve

 

will

 

close

 

just

 

after

 

the

 

shift

 

of

 

the

 

4

way

 

valve.

If

 

main

 

power

 

is

 

lost,

 

the

 

Battery

 

Free

 

Power

 

Loss

 

Alarm

 

will

 

sound

 

a

 

short

 ʺ

BEEP

ʺ

,

 

with

 

a

 

long

 

pause

 

after.

 

All

 

units

 

are

 

equipped

 

with

 

a

 

diagnostic

 

alarm

 

system

 

that

 

signals

 

if

 

the

 

pneumatic

 

pressure

 

or

 

electrical

 

systems

 

malfunction.

 

The

 

troubleshooting

 

guide

 

in

 

this

 

manual

 

explains

 

the

 

alarm

 

system

 

signals

 

and

 

reasons,

 

in

 

detail.

 

Refer

 

to

 

Troubleshooting

 

on

 

page 16.

SensO

2

 Oxygen Sensor Technology - Ceramic Zirconia 

Sensor

Technical Description

The

 

oxygen

 

being

 

produced

 

by

 

the

 

concentrator

 

flows

 

out

 

of

 

the

 

product

 

tank

 

and

 

into

 

the

 

flowmeter.

 

A

 

small

 

flow

 

of

 

oxygen

 

produced

 

by

 

the

 

unit

 

is

 

sent

 

through

 

a

 

precision

 

orifice

 

to

 

the

 

oxygen

 

sensor

 

mounted

 

on

 

the

 

printed

 

circuit

 

board.

As

 

the

 

oxygen

 

enters

 

the

 

sensor,

 

it

 

passes

 

through

 

a

 

screen

 

and

 

contacts

 

the

 

sensing

 

disk.

Electric

 

current

 

flowing

 

through

 

a

 

metal

 

film

 

resistor

 

heats

 

the

 

disk

 

in

 

excess

 

of

 

300°

 

C.

 

Oxygen

 

molecules

 

contact

 

the

 

electrode

 

of

 

the

 

disk

 

and

 

pick

up

 

extra

 

electrons

 

to

 

become

 

oxygen

 

ions.

 

These

 

oxygen

 

ions

 

are

 

attracted

 

to

 

the

 

electrode

 

on

 

the

 

bottom

 

of

 

the

 

zirconia

 

sensing

 

disk.

 

Because

 

of

 

the

 

crystal

 

structure

 

of

 

the

 

zirconia,

 

only

 

oxygen

 

ions

 

can

 

pass

 

through.

 

When

 

the

 

oxygen

 

ions

 

reach

 

the

 

bottom

 

electrode,

 

the

 

extra

 

electrons

 

are

 

released

 

from

 

the

 

oxygen

 

ions

 

and

 

oxygen

 

molecules

 

return

 

to

 

the

 

air.

 

The

 

number

 

of

 

electrons

 

is

 

directly

 

related

 

to

 

the

 

oxygen

 

concentration.

 

The

 

electrons

 

travel

 

to

 

the

 

P.C.

 

board

 

where

 

they

 

are

 

counted

 

and

 

the

 

oxygen

 

concentration

 

reading

 

is

 

calculated.

A

 

microprocessor

 

on

 

the

 

P.C.

 

board

 

contains

 

software

 

that

 

interprets

 

the

 

signal

 

being

 

received

 

from

 

the

 

sensor.

 

It

 

compares

 

the

 

signal

 

to

 

clinically

 

acceptable

 

limits.

 

Signals

 

outside

 

of

 

the

 

clinically

 

acceptable

 

limits

 

generate

 

responses

 

in

 

the

 

form

 

of

 

lights,

 

audible

 

indicators,

 

and/or

 

system

 

shut

down.

Operating Sequence

Once

 

the

 

power

 

switch

 

has

 

been

 

turned

 

on

 

(

)

,

 

the

 

SensO

2

 

circuit

 

will

 

wait

 

five

 

minutes

 

for

 

the

 

concentrator

 

to

 

begin

 

producing

 

clinically

 

acceptable

 

oxygen

 

and

 

the

 

oxygen

 

sensor

 

to

 

stabilize.

 

The

 

GREEN

 

light

 

will

 

illuminate

 

(indicating

 

normal

 

system

 

operation)

 

while

 

the

 

oxygen

 

sensor

 

is

 

warming

 

up.

I/

Содержание Perfecto2 V IRC5PO2V

Страница 1: ...procedures in this manual MUST be performed by a qualified technician For more information regarding Invacare products parts and services please visit www invacare com Perfecto2 V Oxygen Concentrator...

Страница 2: ...n with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products PERFECTO2 V PARTS COMPATIBILITY The Plat...

Страница 3: ...l Description 13 Operating Sequence 13 SECTION 3 PNEUMATIC DIAGRAM 15 SECTION 4 TROUBLESHOOTING 16 SECTION 5 CABINET 23 Removing Cabinet 23 SECTION 6 PREVENTIVE MAINTENANCE 24 Cleaning the Cabinet Fil...

Страница 4: ...l Panel 53 SECTION 14 COOLING FAN 55 Replacing Cooling Fan 55 SECTION 15 P C BOARD 58 Replacing P C Board 58 SECTION 16 TRANSFORMER 61 Replacing the Transformer 61 SECTION 17 ON OFF SWITCH 62 Replacin...

Страница 5: ...ction Test 78 SECTION 24 ALARM TEST 80 Alarm and Sensor Tests 80 Power Loss Alarm 80 Low Pressure Test 80 Time Out Test 81 High Pressure Test 81 Oxygen Sensor 82 P E Valve Coil Test 84 SECTION 25 SIEV...

Страница 6: ...where oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen e...

Страница 7: ...cessory IRCPF16 HomeFill home oxygen compressor IOH200 NOTE This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor The outlet fitting does not aff...

Страница 8: ...IRCPF16 Accessory Low Flow Alarm 0 L min to 0 5 L min Rapid Audible Alarm Beeping No Accessories Connected Average Power Consumption 325 W Pressure Relief Mechanism Operational at 35 psi 5 psi 241 kPa...

Страница 9: ...or similar surfaces Tubing 7 ft cannula with a maximum 50 ft of Crush Proof Tubing DO NOT pinch Relative Humidity 20 to 60 Time of Operation Up to 24 hours per day Recommended Storage and Shipping Te...

Страница 10: ...t away from the oxygen concentrator tubing and connections and all other oxygen equipment DO NOT use any lubricants unless recommended by Invacare For optimum performance Invacare recommends that each...

Страница 11: ...O NOT reach for product that has fallen into water Unplug IMMEDIATELY A product should NEVER be left unattended when plugged in Use this product for only intended use as described in this manual DO NO...

Страница 12: ...plies 120 VAC to the compressor motor hour meter transformer cooling fan and the printed circuit P C board Room air enters the compressor via the cabinet filter and the compressor inlet filter The air...

Страница 13: ...duced by the unit is sent through a precision orifice to the oxygen sensor mounted on the printed circuit board As the oxygen enters the sensor it passes through a screen and contacts the sensing disk...

Страница 14: ...fore activating an alarm Environmental factors such as low voltage high altitude or age of the machine will affect the time required to reach 85 2 If the oxygen level is not above 85 2 within the firs...

Страница 15: ...tput Check Valve Patient Output Oxygen Sensor SensO2 models Only Pressure Sensor P S I O2 P C Board Pressure Regulator Product Tank Check Valves Air Inlet Cabinet Filter Compressor Inlet Filter Sieve...

Страница 16: ...extension cords Move to another outlet or circuit Power cord a Frayed b Broken or damaged spade c Spade connector from power cord loose or disconnected inside back of unit a Reattach cord b Replace p...

Страница 17: ...One Flash GREEN Two Flashes Unit plugged in power switch on continuous alarm Compressor shut down P C board a Malfunction b Disconnected wire c Shifting valve at pressures greater than 25 p s i for P...

Страница 18: ...ctivates or energizes approximately one second prior to the activation of the 4 way valve with approximately 24 volts If voltage is in excess of 24 volts consistently replace the P C board Refer to Re...

Страница 19: ...or a Defective b Faulty capacitor c Bad motor windings d Worn seals e Bad bearings f Leak at fittings or tubing g Leaky or defective relief valve h Insufficient voltage outlet a Replace compressor Ref...

Страница 20: ...uating Flow Regulator Flowmeter a Incorrectly set regulator b Flowmeter malfunction a Check pressure at oxygen outlet Adjust regulator b If flow is still unstable check for leaks starting at the compr...

Страница 21: ...1 b Replace capacitor Refer to Replacing Capacitor on page 33 c Replace compressor d Replace compressor e Replace compressor Line voltage excessive surge Have line voltage inspected by certified elect...

Страница 22: ...Flashing GREEN On Internal repairs required Replace SensO2 circuit board Refer to Replacing P C Board on page 58 Low Flow Alarm does not activate on flows less than 0 5 L min System leak Repair leak i...

Страница 23: ...ng NOTE For this procedure refer to FIGURE 5 1 1 Unplug unit 2 Remove the four mounting screws that secure cabinet assembly to the base assembly 3 Lift the cabinet straight up NOTE When required vacuu...

Страница 24: ...age 16 for plugged filter symptoms Power should be disconnected before beginning preventive maintenance on the Invacare concentrator Cleaning the Cabinet Filter NOTE For this procedure refer to FIGURE...

Страница 25: ...edure refer to FIGURE 6 2 NOTE Perform this procedure as needed depending on the outlet HEPA filter check results 1 Unplug the unit 2 Remove the cabinet Refer to Removing Cabinet on page 23 3 For easi...

Страница 26: ...sor inlet filter by inserting the filter into the rubber base 6 Push the filter down until the rubber base touches the edge of the filter 7 Reinstall the filter access panel FIGURE 6 3 Replacing the C...

Страница 27: ...in the sound box 7 Cut the two tie wraps around the muffler assembly 8 Separate the muffler assembly Detail A in FIGURE 6 5 9 Inspect the components to ensure they are clear 10 Do one of the following...

Страница 28: ...moving Cabinet on page 23 FIGURE 6 5 Replacing the Muffler Assembly Exhaust Canister Throttling Muffler Sound Box Top of Manifold Exhaust Tube Tie Wraps Rubber Grommet DETAIL A Muffler Grommet Throttl...

Страница 29: ...outlet before servicing NOTE For this procedure refer to FIGURE 6 6 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 23 CAUTION Use care not to deform heat exchanger when installing r...

Страница 30: ...R COMES FIRST Check Oxygen Concentration Clean Replace Cabinet Filter s Check Outlet HEPA Filter Check Compressor Inlet Filter Check Power Loss Alarm AS REQUIRED Clean Heat Exchanger Replace Exhaust M...

Страница 31: ...r wires from the top of the capacitor Detail A 8 Disconnect brass nut from compressor Detail A 9 On the backside of the Perfecto2V loosen the clamp that secures the vinyl tube to the manifold assembly...

Страница 32: ...o 1164934 FIGURE 7 1 Removing Replacing Compressor Assembly Intake Hose Mounting Screws Brass Nut Capacitor DETAIL B DETAIL C Vinyl Tube Clamp Manifold Assembly Mounting Screws Grommets DETAIL A Tie W...

Страница 33: ...edlenose pliers disconnect the spade connectors from the capacitor terminals 4 Cut the tie wrap that secures the capacitor to the base of the concentrator 5 Remove the capacitor from the concentrator...

Страница 34: ...ound scale for head screws eccentric set screws valve flapper screws and pipe plugs T 25 Torx torque wrench for head screws piston cup seal retainer and connecting rod eccentric Two flat blade screwdr...

Страница 35: ...rews NOTE Note placement of the head before removing 3 Carefully remove the compressor head and head seal o ring not shown Valve Plates NOTE For this procedure refer to FIGURE 7 3 NOTE Note the positi...

Страница 36: ...oosen the set screw turn 5 Rotate connecting rod to top dead center 180 and slide the connecting rod eccentric assembly off the shaft through the opening in the housing 6 Secure the rod assembly in a...

Страница 37: ...emoving Compressor and Compressor Heads Rebuilding the Compressor Bottom Intake Valve Flapper Removal Top Exhaust Valve Flapper Removal NOTE For this procedure refer to FIGURE 7 5 1 Remove valve flapp...

Страница 38: ...cedure refer to FIGURE 7 5 on page 37 1 Install the new valve flapper screw square washer and valve flapper NOTE Torque flapper screws to 12 inch lbs Top Exhaust Valve Flapper Installation NOTE For th...

Страница 39: ...e access hole tighten set screw to 40 inch lbs Detail A of FIGURE 7 4 5 Hold the cylinder down to the housing with one hand and slowly rotate the eccentric with the other hand to ensure all the compon...

Страница 40: ...ompressor Refer to Removing Replacing Compressor Assembly on page 31 2 Run unit and check for leaks around compressor hoses and compressor tubes Refer to Leak Test on page 76 3 Take pressure readings...

Страница 41: ...etainer 4 Piston Cup Retainer 5 Cylinder 6 Head 7 Head Screw 8 Valve Plate Assembly 9 Valve Flapper Keeper Strip 10 Square Washer 11 Valve Flapper Top Bottom 12 Valve Plate 13 Valve Flapper Screw 14 F...

Страница 42: ...tie wraps from P E valve tubing 6 Remove P E valve assembly from tubing CAUTION If P E valve is not to be installed IMMEDIATELY plug the P E valve tubing to prevent sieve bed contamination 7 Install...

Страница 43: ...rt No 1164934 43 Perfecto2 V FIGURE 8 1 Replacing P E Valve OUT Tie Wrap P E Valve Tubing P E Valve Tubing P E Valve Assembly Sieve Bed Out Port Barb Tie Wrap P E Valve Terminals Spade Connectors Atta...

Страница 44: ...k and lift up Rest sieve beds on rear of base 7 Using a hose clamp tool remove the clamp and reinforced tubing from bottom of both sieve beds and remove sieve beds CAUTION DO NOT remove plastic caps f...

Страница 45: ...Wrap P E Valve Tubing Tie Wraps P E Valve Assembly Reinforced Tubing attaches HERE not shown NOTE Compressor Inlet HEPA Filter removed from illustration for clarity Tie Wrap P E Valve Tubing Tie Wrap...

Страница 46: ...correctly The letters VAC are printed on the check valves The check valves MUST be installed with the VAC on the underside of the check valve Low system pressures and eventual bed contamination will...

Страница 47: ...PA Filter on page 26 9 Run units and check for leaks Refer to Leak Test on page 76 10 Reinstall cabinet Refer to Removing Cabinet on page 23 FIGURE 10 1 Replacing Check Valves Tie wrap Check Valves Pr...

Страница 48: ...moved NOTE Clean mounting surface of product tank cap Ensure that the two O rings for the NEW regulator are installed onto the ports underside of the NEW regulator before installation WARNING Ensure p...

Страница 49: ...5 Plug unit in and turn power switch on 6 Using the access hole located on the control panel locate the pressure adjustment screw in center of pressure regulator top 7 Insert a 5 32 inch Allen wrench...

Страница 50: ...10 Once required pressure is achieved reinstall cabinet Refer to Removing Cabinet on page 23 11 Refer to Troubleshooting on page 16 if you are unable to adjust or maintain five p s i 0 5 p s i NOTE Af...

Страница 51: ...emove cabinet Refer to Removing Cabinet on page 23 CAUTION Use care not to deform heat exchanger when installing removing or cleaning 3 Disconnect brass nut that secures the heat exchanger assembly to...

Страница 52: ...EXCHANGER ASSEMBLY Perfecto2 V 52 Part No 1164934 FIGURE 12 1 Replacing Heat Exchanger Assembly Compressor Sound Box Heat Exchanger Assembly DETAIL A DETAIL B Brass Nut Clamp Vinyl Tube Manifold Asse...

Страница 53: ...o the sound box Detail B 6 Remove the inch I D tubing from the top and bottom barbed fittings of the flowmeter Detail B 7 Cut tie wrap and remove inch I D tubing from the patient outlet barbed fitting...

Страница 54: ...p Refer to Leak Test on page 76 16 Reinstall cabinet Refer to Removing Cabinet on page 23 FIGURE 13 1 Removing Control Panel Circuit Breaker STEP 9 Hour Meter STEP 9 On Off Switch STEP 9 Mounting Scre...

Страница 55: ...ecessary Detail A 7 Disconnect spade connectors from fan terminals on side of cooling fan NOTE The cooling fan is fastened to the soundbox with four rubber grommets on each corner 8 Lift cooling fan u...

Страница 56: ...he bottom of the housing Detail A 18 if necessary reconnect the intake valve on the compressor assembly 19 Reinstall control panel Refer to Removing Control Panel on page 53 20 Reinstall compressor in...

Страница 57: ...Compressor Assembly DETAIL A DETAIL B Fan Terminals Spade Connectors attach here Cooling Fan Rubber Grommet DETAIL C Cooling Fan Rubber Grommets Soundbox DETAIL D NOTE Cooling fan removed from rubber...

Страница 58: ...n upward force with a slight rocking motion to remove the terminals Before installing any P C boards ensure that all insulators are in place NOTE For this procedure refer to FIGURE 15 1 on page 60 1 U...

Страница 59: ...o product tank C For concentrators that have O2 sensor Connect the oxygen sensor tubing to regulator fitting 9 Reinstall control panel Refer to Removing Control Panel on page 53 10 Run concentrator to...

Страница 60: ...C BOARD Perfecto2 V 60 Part No 1164934 FIGURE 15 1 Replacing P C Board Control Panel Mounting Screw P C Board Oxygen Sensor Tubing IRC5PO2 only Pressure Sensor Tubing Mounting Screw Wiring Harnesses...

Страница 61: ...r located on the sound box 4 Code all wires before disconnecting for proper re connections 5 Remove the two mounting screws that secure the transformer to the soundbox 6 Note orientation of transforme...

Страница 62: ...NOTE Ensure proper orientation of the On Off switch and wiring connections before installation Refer to FIGURE 17 1 Detail A and Detail B below CAUTION DO NOT install the On Off I O switch upside down...

Страница 63: ...w flowmeter is installed 4 Remove control panel Refer to Removing Control Panel on page 53 5 Remove palnuts that secure flowmeter to the control panel 6 Remove flowmeter from front of control panel 7...

Страница 64: ...53 4 Remove and label the two hour meter connectors and wires connected to back of hour meter 5 Spread retaining clips on hour meter housing that secure hour meter to control panel 6 Remove hour meter...

Страница 65: ...the 4 way valve manifold assembly 6 Remove the F tube from the 4 way valve manifold assembly 7 Using a hose clamp tool remove the clamps that secure the vinyl tubes to the 4 way valve manifold assemb...

Страница 66: ...e side of the 4 way valve assembly 19 Position the 4 way valve manifold assembly in the base of the concentrator on the foam block in the orientation noted in STEP 9 20 Slide the vinyl tubes onto each...

Страница 67: ...NOTE The Numbers above indicate the MOUNTING SCREW TORQUE SEQUENCE pre torque to 10 2 inch lbs then torque to 22 2 inch lbs Mounting Screws 4 Way Valve Manifold Assembly Manifold Exhaust Tube Tie Wra...

Страница 68: ...stem 7 Remove pilot valve poppet from inside the pilot valve stem CAUTION DO NOT remove the coil yoke from the coil The washer between the bottom of the yoke and the bottom of the coil MUST not be re...

Страница 69: ...coil with yoke at an approximate 90 angle to the right as viewed from the front of valve 17 Use the flat blade screwdriver and hold the pilot valve stem in place 18 Tighten locknut clockwise with the...

Страница 70: ...lve Poppets and O Rings Spade Connectors attach here Leave Wire Connectors Intact Pilot Valve Stem Yoke New Label Locknut Coil Pilot Valve Stem Valve Manifold Tapered End O Ring Slotted Poppet NOTE DO...

Страница 71: ...rity to reach maximum When the unit is turned on the GREEN light will come on SYSTEM OK O2 greater than 85 After five minutes the oxygen sensor will be operating normally and will control the indicato...

Страница 72: ...SECTION 21 PURITY INDICATORS Perfecto2 V 72 Part No 1164934 FIGURE 21 1 Display Control Panel Auto Shut Down Below Normal Normal Model IRC5P02V...

Страница 73: ...or this procedure refer to FIGURE 22 1 on page 74 1 Turn power off and unplug the unit 2 Remove cabinet Refer to Removing Cabinet on page 23 3 Plug the unit in and turn the power on 4 Locate the P C b...

Страница 74: ...sfully completed before the unit can be placed into service 12 After a successful autotune run the unit for 30 minutes and verify that the oxygen concentration output of the unit is within specificati...

Страница 75: ...5 minutes of run time the concentration is below specification change setting to a number one position lower than the previous setting 9 Let unit run ten minutes If concentration is lower still change...

Страница 76: ...es at 4 way valve DO NOT allow leak test solution to enter into 4 way valve and system 6 With leak test solution or equivalent soapy water leak test the following A The two sieve bed hose connection a...

Страница 77: ...Valve Manifold Assembly STEP 6B Sieve Bed Hose Connection At 4 way Valve manifold Assembly STEP 6A 4 Way Valve Manifold Assembly Hose Connections to Bottom Fittings of Sieve Beds STEP 6C Hose Connect...

Страница 78: ...e These leaks are detected by either reading the system swing pressures noticing two consecutive swings not being equal or by hearing an audible leak every other cycle Leaks of this extreme nature ind...

Страница 79: ...Function Test 4 Way Valve Check each end cap for cracks and for major leaks You should NOT use leak detector on the valve Use a 0 30 psi pressure gauge to monitor system pressure swings and also recor...

Страница 80: ...in and switched on Low Pressure Test NOTE For this procedure refer to FIGURE 24 1 on page 81 1 There are two separate failure modes for Low Pressure Low product tank pressure Pressure in tank drops b...

Страница 81: ...Time Out Test High Pressure Test NOTE For this procedure refer to FIGURE 24 2 on page 82 High Pressure Occurs when pressure in product tank rises beyond a preset value typically 23 p s i 1 p s i for 5...

Страница 82: ...oncentration levels fall below a preset value typically 73 3 or 85 2 This failure indicates a catastrophic failure of any component mechanical or electrical 1 Turn unit on Set output flow at 5 L min 2...

Страница 83: ...en clamping the oxygen sensor tubing 6 Within 30 minutes the Oxygen Sensor Alarm should activate and the RED panel light indicator will illuminate The compressor will shut down with the alarm FIGURE 2...

Страница 84: ...e coil alarm should activate within 10 seconds 2 Replace YELLOW wire and reinstall cabinet Refer to Removing Cabinet on page 23 NOTE If any alarm fails to perform to specification contact Invacare Tec...

Страница 85: ...SECTION 24 ALARM TEST Part No 1164934 85 Perfecto2 V FIGURE 24 1 Alarm Diagnostic Label...

Страница 86: ...Turn power off and unplug unit 2 Turn unit so the rear of the unit is facing you FIGURE 25 1 Post Pressure Check Pressure Testing NOTE For this procedure refer to FIGURE 25 2 on page 87 1 Connect pre...

Страница 87: ...hift the pressure in the concentrator will drop to 12 p s i g 2 before rising within 20 seconds to 21 p s i g 0 3 and shift the 4 way valve to the opposite sieve bed 8 If pressure reading is 21 p s i...

Страница 88: ...ssissauga Ontario L4Z 4G4 Canada 800 668 5324 2010 Invacare Corporation All rights reserved Republication duplication or modification in whole or in part is prohibited without prior written permission...

Отзывы: